Assessment of two commercial agglutination kits for detecting Escherichia coli heat-labile enterotoxin.

Can J Microbiol

Bureau of Microbial Hazards, Health and Welfare Canada, Ottawa, Ont.

Published: November 1991

Two commercial agglutination kits, a reserved passive agglutination test (VET-RPLA) and a staphylococcal coagglutination test (Phadebact ETEC-LT Test), were compared with two cell culture assays (Y-1 and Vero) and GM1 ganglioside enzyme-linked immunosorbent assay (GM1-ELISA) for sensitivity in detecting Escherichia coli heat-labile enterotoxin (LT). Of 48 toxigenic strains, 23 were positive by all assays. One strain was negative only by the Phadebact test. Four strains, all LT-II producers, were positive by cell culture only. For LT-I detection, the Phadebact test was the least sensitive but was simple and rapid; VET-RPLA was simple, sensitive, and a good substitute for cell culture or GM1-ELISA.

Download full-text PDF

Source
http://dx.doi.org/10.1139/m91-151DOI Listing

Publication Analysis

Top Keywords

cell culture
12
commercial agglutination
8
agglutination kits
8
detecting escherichia
8
escherichia coli
8
coli heat-labile
8
heat-labile enterotoxin
8
phadebact test
8
test
5
assessment commercial
4

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?