Accuracy of bedside capillary blood glucose measurements in critically ill patients.

Intensive Care Med

Department of Medicine, Thomas Jefferson University, 1025 Walnut Street, Suite 821, Philadelphia, PA 19107, USA.

Published: December 2007

Objective: To compare the accuracy of fingerstick with laboratory venous plasma glucose measurements (laboratory glucose) in medical ICU patients and to determine the factors which interfere with the accuracy of fingerstick measurements.

Participants: The study included 80 consecutive patients aged 58+/-7 years, BMI 29.5+/-9.0, and APACHE II score 15+/-6 (277 simultaneous paired measurements).

Measurements: This prospective observational study compared fingerstick measurements to simultaneously sampled laboratory glucose once a day in patients in our medical ICU (twice daily if on an insulin infusion). Data recorded included patient demographics, admission diagnoses, APACHE II score, BMI, daily hematocrit, arterial blood gasses, chemistry results, concomitant medications (including vasopressors and corticosteroids), and upper extremity edema. Accuracy was defined as the percentage of paired values not in accord (>15 mg dl(-1)/ 0.83 mmol(-1)l(-1) difference for laboratory values <75 mg dl(-1)/4.12 mmol(-1)l(-1) and >20% difference for laboratory values >or=75 mg/dl). Outliers (blood glucose difference >100 mg dl(-1)/5.56 mmol(-1) l(-1)) were excluded from the correlation and distribution analyses.

Results: Mean fingerstick glucose was 129+/-45 mg/dl (7.2+/-2.5 mmol/l) and mean laboratory glucose 123+/-44 mg/dl (6.8+/-2.4 mmol/l). The correlation coefficient between the two values was 0.9110 (Clinical and Laboratory Standards Institute threshold 0.9751). The mean difference (bias) between the two methods was 8.6+/-18.6 mg/dl (0.48+/-1.0 mmol/l) and limits of agreement +45.8 and -28.6 mg/dl (+2.5 and -1.6 mmol/l). Fifty-three (19%) paired measurements in 22 patients were not in accord (CLSI threshold
Conclusions: The findings suggest that capillary blood glucose as measured by fingerstick is inaccurate in critically ill ICU patients and does not meet the CLSI standard. It is unclear whether the sampling method, device used, or both contributed to this inaccuracy. The wide limits of agreement suggest that fingerstick measurements should be used with great caution in protocols of tight glycemic control.

Download full-text PDF

Source
http://dx.doi.org/10.1007/s00134-007-0835-4DOI Listing

Publication Analysis

Top Keywords

laboratory glucose
12
blood glucose
8
glucose measurements
8
accuracy fingerstick
8
medical icu
8
apache score
8
difference laboratory
8
laboratory values
8
glucose
7
laboratory
7

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!