Fenamidone is an imidazolinone fungicide recently introduced in viticulture practices. This work reports the validation and assessment of global uncertainty of a gas chromatographic with mass spectrometry method to analyze fenamidone in grapes and wines. This method consists in a simple and fast liquid-liquid extraction step followed by chromatographic determination. Limits of detection for fenamidone in grapes and wines were, respectively, 0.05 mg/kg and 0.06 mg/L, precision was below 9.4% and average recovery was 89 +/- 5%. In the concentration range from 0.05 to 1.00 mg/kg (or mg/L) of fenamidone, global uncertainty calculated following the EURACHEM/CITAC rules, and also by the Horwitz function, was below 25%. The EURACHEM/CITAC global uncertainty budget used gave lower estimates than those obtained from the Horwitz function.