Measuring key parameters of intense pulsed light (IPL) devices.

J Cosmet Laser Ther

Independent Laser Protection Adviser, Haywards Heath, West Sussex, UK.

Published: September 2007

Background: Unlike medical lasers, intense pulsed light (IPL) devices are largely unregulated and unclassified as to degree of safety hazard. With the exception of most of the USA, the United Kingdom and parts of Europe, the Far East and Australia, the sale of IPLs is generally unrestricted, with the majority being sold into the beauty therapy and spa markets. Standards are only imposed on manufacturers for technical performance data and operating tolerances determined by CE-compliance under electrical safety standards or the EU Medical Device Directive. Currently, there is no requirement for measurement of key IPL performance characteristics.

Objective: To identify the key IPL parameters, emphasize their importance in terms of safe and effective treatment and provide examples of preliminary measurement methods. These measurements can highlight changes in an IPL device's performance, improving patient safety and treatment efficacy.

Methods: Five key parameters were identified as having an important role to play in the way light interacts with the skin, and therefore an important role in patient safety and effective treatment. Simple methods were devised to measure the parameters, which include fluence, pulse duration, pulse profile, spectral output and time-resolved spectral output.

Results: The measurement methods permitted consistent and comparable measurements to be made by two of the authors at working clinic locations on 18 popular IPL devices and allowed assessment of output variations. Results showed discrepancies between the measured IPL device outputs and those values displayed on the system or claimed by the manufacturers. The importance of these discrepancies and their impact is discussed.

Conclusions: This study, of 18 popular devices in regular daily use in England and Wales, provides example methods for measuring key IPL device parameters and highlights the need for regular measurement of at least those five key parameters measured in this study. These methods can help service technicians to check performance and eliminate device malfunction.

Download full-text PDF

Source
http://dx.doi.org/10.1080/14764170701435297DOI Listing

Publication Analysis

Top Keywords

key parameters
12
ipl devices
12
key ipl
12
measuring key
8
intense pulsed
8
pulsed light
8
ipl
8
light ipl
8
measurement key
8
effective treatment
8

Similar Publications

Objective: Using F-FDG PET/CT metabolic parameters to differentiate post-transplant lymphoproliferative disorder (PTLD) and reactive lymphoid hyperplasia (RLH), and PTLD subtypes.

Methods: F-FDG PET/CT and clinical data from 63 PTLD cases and 19 RLH cases were retrospectively collected. According to the 2017 WHO classification, PTLD was categorized into four subtypes: nondestructive (ND-PTLD), polymorphic (P-PTLD), monomorphic (M-PTLD), and classic Hodgkin.

View Article and Find Full Text PDF

Development of Electrospinning Setup for Vascular Tissue-Engineering Application with Thick-Hierarchical Fiber Alignment.

Tissue Eng Regen Med

January 2025

College of Materials Science and Engineering, Hunan University, Changsha, 410072, People's Republic of China.

Background: Tissue engineering holds promise for vascular repair and regeneration by mimicking the extracellular matrix of blood vessels. However, achieving a functional and thick vascular wall with aligned fiber architecture by electrospinning remains a significant challenge.

Methods: A novel electrospinning setup was developed that utilizes an auxiliary electrode and a spring.

View Article and Find Full Text PDF

Background: Spinocerebellar ataxia type 3 (SCA3) is a neurodegenerative disorder, with balance instability as a feature of the disease. Balance instability often manifests before the onset of obvious ataxic symptoms in patients. However, current clinical scales exhibit limited sensitivity in characterizing changes in pre-ataxic patients.

View Article and Find Full Text PDF

Objective: Soft tissue defects and postoperative wound healing complications related to calcaneus fractures may result in significant morbidity. The aim of this study was to investigate whether percutaneous minimally invasive screw internal fixation (PMISIF) can change this situation in the treatment of calcaneal fractures, and aimed to explore the mechanical effects of different internal fixation methods on Sanders type III calcaneal fractures through finite element analysis.

Methods: This retrospective analysis focused on 83 patients with Sanders II and III calcaneal fractures from March 2017 to March 2022.

View Article and Find Full Text PDF

With the global rise in advanced maternal age (AMA) pregnancies, the risk of gestational diabetes mellitus (GDM) increases. However, few GDM prediction models are tailored for AMA women. This study aims to develop a practical risk prediction model for GDM in AMA women.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!