Background: Glucose testing at the bedside has become an integral part of the management strategy in diabetes and of the careful maintenance of normoglycemia in all patients in intensive care units. We evaluated two point-of-care glucometers for the determination of plasma-equivalent blood glucose.

Methods: The Precision PCx and the Accu-Chek Inform glucometers were evaluated. Imprecision and bias relative to the Vitros 950 system were determined using protocols of the Clinical Laboratory Standards Institute (CLSI). The effects of low, normal, and high hematocrit levels were investigated. Interference by maltose was also studied.

Results: Within-run precision for both instruments ranged from 2-5%. Total imprecision was less than 5% except for the Accu-Chek Inform at the low level (2.9 mmol/L). Both instruments correlated well with the comparison instrument and showed excellent recovery and linearity. Both systems reported at least 95% of their values within zone A of the Clarke Error Grid, and both fulfilled the CLSI quality criteria. The more stringent goals of the American Diabetes Association, however, were not reached. Both systems showed negative bias at high hematocrit levels. Maltose interfered with the glucose measurements on the Accu-Chek Inform but not on the Precision PCx.

Conclusions: Both systems showed satisfactory imprecision and were reliable in reporting plasma-equivalent glucose concentrations. The most stringent performance goals were however not met.

Download full-text PDF

Source
http://dx.doi.org/10.1016/j.cca.2007.07.025DOI Listing

Publication Analysis

Top Keywords

accu-chek inform
16
plasma-equivalent glucose
8
precision pcx
8
high hematocrit
8
hematocrit levels
8
glucose point-of-care
4
point-of-care evaluation
4
evaluation roche
4
accu-chek
4
roche accu-chek
4

Similar Publications

This study assessed the reliability of Roche Accu-Chek Inform II glucometers in a real-world setting. A retrospective analysis was conducted on 6,695 paired results. Capillary samples were tested using Roche Accu-Chek Inform II glucometers, while venous samples were analyzed using Roche Cobas c503/702 analyzers.

View Article and Find Full Text PDF

Background: The need for frequent blood glucose (BG) monitoring and the inconveniences associated with self-monitoring of BG (SMBG) have driven the development of non-invasive approaches.

Methods: This prospective study aimed to investigate the accuracy of glucose level calculation using the near-infrared spectroscopy (NIRS) technology system. People with Type 1 diabetes, Type 2 diabetes, prediabetes, and normal glucose metabolism were included.

View Article and Find Full Text PDF

Predicting Glucose Values: A New Era for Continuous Glucose Monitoring.

J Diabetes Sci Technol

September 2024

Science Consulting in Diabetes GmbH, Düsseldorf, Germany.

The last 25 years of CGM have been characterized above all by providing better and more accurate glucose values in real time and analyzing the measured glucose values. Trend arrows are the only way to look into the future, but they are often too imprecise for therapy adjustment. While AID systems provide algorithms to use glucose values for glucose control, this has not been possible with stand-alone CGM systems, which are most used by people with diabetes.

View Article and Find Full Text PDF

Enhancing the Capabilities of Continuous Glucose Monitoring With a Predictive App.

J Diabetes Sci Technol

September 2024

Roche Diabetes Care Deutschland GmbH, Mannheim, Germany.

Background: Despite abundant evidence demonstrating the benefits of continuous glucose monitoring (CGM) in diabetes management, a significant proportion of people using this technology still struggle to achieve glycemic targets. To address this challenge, we propose the Accu-Chek SmartGuide Predict app, an innovative CGM digital companion that incorporates a suite of advanced glucose predictive functionalities aiming to inform users earlier about acute glycemic situations.

Methods: The app's functionalities, powered by three machine learning models, include a two-hour glucose forecast, a 30-minute low glucose detection, and a nighttime low glucose prediction for bedtime interventions.

View Article and Find Full Text PDF

Aims/introduction: Blood glucose meters are commonly used at the bedside, but most of the meters used in Hung Vuong Hospital (Ho Chi Minh City, Vietnam) are built for self-monitoring and might not be suitable for determining glucose levels in patients. In this study, we aimed to validate the performance of six frequently used meters in our hospital using the Clinical & Laboratory Standards Institute (CLSI) standard, and investigate the hematocrit impact on the accuracy of these meters.

Materials And Methods: A total of 135 pregnant women who underwent a 75-g oral glucose tolerance test consented to participate in the study at Hung Vuong Hospital.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!