Three-dimensional numerical simulations of multi-cryo-needle surgery were performed with cryo-needle temperature variations taken from matched experimental data. The transient temperatures and frozen volumes generated by simultaneously operating up to three 1.47 mm OD cryo-needles embedded in a phase-changing gel simulating the properties of biological tissues, were studied. In all cases studied, the volumes enclosed by the "lethal", -40 degrees C isotherm, achieved most of their final size in the first few minutes of operation, thus obviating the need for long application times. After 30 min of application of the one-, two- or three-cryo-needles, the ablation ratio attained 3%, 3-6% and 3-8%, respectively, depending on cryo-needle placement configurations. Synergistic effects of using multi-cryo-needles were reflected in the increased expansion of both the radial and axial locations of the isothermal contours. Within each number of cryo-needles used, however, the differences in these locations were rather small, and, as a general rule, tended to somewhat decrease with increasing the placement "density" of the cryo-needles. For each two- and three-cryo-needle application, there is a certain combination of placement configuration and application time that would produce the largest, temperature-specific, volume. As a general guideline, multiple cryo-needles should not be placed too close to each other in order to enhance their synergistic effect. Results of this study should be useful in the design of cryo-needle placement and operation protocols and in understanding the limitations of the freezing-ablation process.

Download full-text PDF

Source
http://dx.doi.org/10.1016/j.cryobiol.2007.06.009DOI Listing

Publication Analysis

Top Keywords

cryo-needle placement
8
cryo-needles
5
isothermal volume
4
volume contours
4
contours generated
4
generated freezing
4
freezing gel
4
gel embedded
4
embedded cryo-needles
4
cryo-needles applications
4

Similar Publications

AI-Based Isotherm Prediction for Focal Cryoablation of Prostate Cancer.

Acad Radiol

September 2023

Brigham and Women's Hospital, 75 Francis St, Boston, MA 22115 (P.M., K.T., C.T., J.T.); Harvard Medical School, 25 Shattuck St, Boston, MA 02115 (P.M., K.T., C.T., J.T.).

Rationale And Objectives: Focal therapies have emerged as minimally invasive alternatives for patients with localized low-risk prostate cancer (PCa) and those with postradiation recurrence. Among the available focal treatment methods for PCa, cryoablation offers several technical advantages, including the visibility of the boundaries of frozen tissue on the intraprocedural images, access to anterior lesions, and the proven ability to treat postradiation recurrence. However, predicting the final volume of the frozen tissue is challenging as it depends on several patient-specific factors, such as proximity to heat sources and thermal properties of the prostatic tissue.

View Article and Find Full Text PDF

Partial gland cryoablation (PGC) aims at destroying prostate cancer (PCa) foci while sparing the unaffected prostate tissue and the functionally relevant structures around the prostate. Magnetic Resonance Imaging (MRI) has boosted PGC, but available evidence suggests that ablation margins may be positive due to MRI-invisible lesions. This study aimed at determining the potential role of intraoperative digital analysis of ablation margins (DAAM) by fluoresce confocal microscopy (FCM) of biopsy cores taken during prostate PGC.

View Article and Find Full Text PDF

The Impact of Placement Errors on the Tumor Coverage in MRI-Guided Focal Cryoablation of Prostate Cancer.

Acad Radiol

June 2021

Department of Radiology, Brigham and Women's Hospital, Harvard Medical School, 75 Francis St. Boston, 02115 Massachusetts, USA.

Article Synopsis
  • This study investigates the accuracy of cryo-needle placement in focal cryoablation treatments for prostate cancer, highlighting its importance for effective tumor ablation.
  • It uses Monte Carlo simulations based on MRI data from 15 patients to model how placement errors affect tumor coverage and the chances of residual cancer.
  • Findings show that, even with placement errors of up to 3 mm, the coverage remains high with two or more needles, but the success rates vary significantly based on the temperature targets used during the procedure.
View Article and Find Full Text PDF

Modeling Cryotherapy Ice Ball Dimensions and Isotherms in a Novel Gel-based Model to Determine Optimal Cryo-needle Configurations and Settings for Potential Use in Clinical Practice.

Urology

May 2016

Division of Surgery and Interventional Science, University College London, London, UK; Department of Urology, University College London Hospitals NHS Foundation Trust, London, UK; Princess Alexandra Hospital NHS Trust, Harlow, UK.

Objective: To gain a better understanding of ice ball dimensions and temperature isotherms relevant for cell kill when using combinations of cryo-needles we set out to answer 4 questions: (1) what type of cryo-needle? (2) how many needles? (3) best spatial configuration? and (4) correct duty cycle percentage?

Methods: We conducted laboratory experiments to monitor ice ball dimensions and create multi-needle planar isotherm maps for 17G and 10G cryo-needles using a novel multi-needle thermocouple fixture within gel at body temperature. We tested configurations of 1-4 cryo-needles at duty cycles of 20%-100% with 1-2.5 cm spacing.

View Article and Find Full Text PDF

Three-dimensional numerical simulations of multi-cryo-needle surgery were performed with cryo-needle temperature variations taken from matched experimental data. The transient temperatures and frozen volumes generated by simultaneously operating up to three 1.47 mm OD cryo-needles embedded in a phase-changing gel simulating the properties of biological tissues, were studied.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!