Objective: This study compared the performance of forehead reflectance and conventional pulse oximetry (SpO(2)) in mechanically ventilated, anesthetized patients undergoing major abdominal surgery with either laparoscopic or laparotomic techniques.
Methods: SpO(2) was continuously measured both with a forehead reflectance and a conventional digit sensor in 20 ASA physical status I-III, anesthetized patients undergoing either a laparotomic (group Laparotomy, n = 10) or laparoscopic (group Laparoscopy, n = 10) major abdominal surgery. SpO(2) values measured with the two sensors were continuously recorded at 10-second intervals during the entire procedure, and then analyzed for consistency.
Results: In group Laparotomy finger and forehead readings remained consistently similar during the study period; on the contrary, in group Laparoscopy forehead SpO(2) readings showed a much wider variability. The mean (95% Confidence Intervals) difference between finger and forehead SpO(2) readings was 2.0% (-1.3% to +6.0%) in group Laparotomy and 2.5% (-3.5% to +8.4%) in group Laparoscopy (p = 0.001); however, in group Laparoscopy the mean difference between digit and forehead SpO(2) values increased from 0.2% (CI(95): -2.1% to +2.5%) during the first hour to 4.5% (CI(95): -2.5% to +11-5%) in the second hour (p = 0.0005), and 3.1% (CI(95): -2.4% to + 8.6%) in the third hour of surgery (p = 0.0005). Clinically relevant desaturation (decrease of SpO(2) < 89% for > or =30 s) detected with the forehead sensor in the Laparoscopy group was significantly more frequent and longer lasting than with conventional digit sensor. No differences were observed in group Laparotomy.
Conclusions: Forehead reflectance oximetry is as accurate as conventional digit based oximetry in mechanically ventilated patients undergoing laparotomic surgery in the supine position, but is significantly influenced by patient positioning and pneumoperitoneum during laparoscopic surgery.
Download full-text PDF |
Source |
---|---|
http://dx.doi.org/10.1007/s10877-007-9084-x | DOI Listing |
Sensors (Basel)
December 2024
Division of Neurological Rehabilitiation, Instituto Nacional de Rehabilitacion Luis Guillermo Ibarra Ibarra, Mexico City 14389, Mexico.
Stroke is a global health issue caused by reduced blood flow to the brain, which leads to severe motor disabilities. Measuring oxygen levels in the brain tissue is crucial for understanding the severity and evolution of stroke. While CT or fMRI scans are preferred for confirming a stroke due to their high sensitivity, Near-Infrared Spectroscopy (NIRS)-based systems could be an alternative for monitoring stroke evolution.
View Article and Find Full Text PDFPediatr Res
October 2024
Optics and Photonics Research Group and Centre for Healthcare Technologies, University of Nottingham, University Park, Nottingham, UK.
J Plast Reconstr Aesthet Surg
November 2024
Department of Cosmetic & Plastic Surgery, Affiliated Hangzhou First People's Hospital, Westlake University Medical College, Hangzhou, China. Electronic address:
Physiol Rep
September 2024
PepsiCo R&D Life Sciences, Gatorade Sports Science Institute, Valhalla, New York, USA.
The purpose of this study was to determine the effect of hydration status on the change in sweat sodium (Na), chloride (Cl), and potassium (K) concentrations during exercise-heat stress. Fifteen subjects (Six female, nine male; 29 ± 9 y; 71 ± 14 kg) completed 90 min of cycling (81% HR) in the heat (~33°C, 42% rh) with fluid replacement to maintain euhydration (EUH) or without fluid to dehydrate to 2.4 ± 0.
View Article and Find Full Text PDFCrit Care
September 2024
AP-HP, Service de médecine intensive-réanimation, Hôpital de Bicêtre, DMU 4 CORREVE, Inserm UMR S_999, FHU SEPSIS, CARMAS, Université Paris-Saclay, 78 rue du Général Leclerc, 94270, Le Kremlin-Bicêtre, France.
Background: To detect preload responsiveness in patients ventilated with a tidal volume (Vt) at 6 mL/kg of predicted body weight (PBW), the Vt-challenge consists in increasing Vt from 6 to 8 mL/kg PBW and measuring the increase in pulse pressure variation (PPV). However, this requires an arterial catheter. The perfusion index (PI), which reflects the amplitude of the photoplethysmographic signal, may reflect stroke volume and its respiratory variation (pleth variability index, PVI) may estimate PPV.
View Article and Find Full Text PDFEnter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!