The femoral components of the total knee replacements are generally made of metal. In contrast, ceramic femoral components promise improved tribological and allergological properties. However, ceramic components present a risk of failure as a result of stress peaks. Stress peaks can be minimised through adequate implant design, proper material composition and optimum force transmission between bone and implant. Thus, the quality of the implant fixation is a crucial factor. The objective of the present study was to analyse the influence of the cement layer thickness on stress states in the ceramic femoral component and in the femur. Two- and three- dimensional finite element analyses of an artificial knee joint with cement layers of different thickness and with an unbalanced cement layer thickness between the ceramic femoral component and the femur were performed. Higher stress regions occurred in the area of force transmission and in the median plane. The maximum calculated stresses were below the accepted tensile strength. Stresses were found to be lower for cement layer thickness of <2.0 mm.