The tissue response to nitric oxide (NO)-releasing subcutaneous implants is presented. Model implants were created by coating silicone elastomer with diazeniumdiolate-modified xerogel polymers capable of releasing NO. The host tissue response to such implants was evaluated at 1, 3, and 6 weeks and compared to that of uncoated silicone elastomer blanks and xerogel-coated controls incapable of releasing NO. Delivery of NO (approximately 1.35 micromol/cm2 of implant surface area) reduced foreign body collagen capsule ("scar tissue") thickness by >50% compared to uncoated silicone elastomer after 3 weeks. The chronic inflammatory response at the tissue/implant interface was also reduced by >30% at NO-releasing implants after 3 and 6 weeks. Additionally, CD-31 immunohistochemical staining revealed approximately 77% more blood vessels in proximity to NO-releasing implants after 1 week compared to controls. These findings suggest that conferring NO release to subcutaneous implants may promote effective device integration into healthy vascularized tissue, diminish foreign body capsule formation, and improve the performance of indwelling medical devices that require constant mass transport of analytes (e.g., implantable sensors).

Download full-text PDF

Source
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC2692676PMC
http://dx.doi.org/10.1016/j.biomaterials.2007.06.036DOI Listing

Publication Analysis

Top Keywords

foreign body
12
subcutaneous implants
12
silicone elastomer
12
reduced foreign
8
response nitric
8
tissue response
8
compared uncoated
8
uncoated silicone
8
no-releasing implants
8
implants
7

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!