Fourier transform capillary electrophoresis (FTCE) was developed as a method to improve signal-to-noise ratio (S/N) and resolution in capillary electrophoresis (CE) separation. In FTCE, multiple simultaneous CE separations were performed in the same channel system and interrogated using a single-point detector. To illustrate experimentally the improvement offered by FTCE in S/N ratio and resolution, we carried out a modest number (five) of multiple injections and separations. We show even with this small number of separations, S/N increased by a factor of 2.9, and theoretical plate height improved by a factor of more than 30. We demonstrated this technique with laser-induced fluorescence detection, but a wide variety of detection methods are compatible with FTCE.
Download full-text PDF |
Source |
---|---|
http://dx.doi.org/10.1021/ac0710026 | DOI Listing |