34-year durability of a DeBakey Surgitool mechanical aortic valve prosthesis.

Interact Cardiovasc Thorac Surg

Department of Cardiac Surgery, University Hospital Grosshadern, Ludwig-Maximilians-University, Marchioninistrasse 15, 81377 Munich, Germany.

Published: October 2006

Background: The DeBakey Surgitool mechanical aortic valve prosthesis was the first mechanical prosthetic valve in which pyrolitic carbon was used. This valve was on the market from 1969 to 1978 and approximately 3300 valves were implanted in this lapse worldwide.

Report: We present a DeBakey Surgitool mechanical aortic valve prosthesis implanted in 1972, which has been functioning for almost 34 years and still presents a good performance.

Comment: Spontaneous strut fractures associated with haemodynamic complications and death, even in asymptomatic patients, was the reason to withdraw this prosthetic valve from the market. However, a prophylactic replacement of DeBakey Surgitool valve was not indicated in our patient due to the excellent prosthesis compliance.

Download full-text PDF

Source
http://dx.doi.org/10.1510/icvts.2006.135087DOI Listing

Publication Analysis

Top Keywords

debakey surgitool
16
surgitool mechanical
12
mechanical aortic
12
aortic valve
12
valve prosthesis
12
prosthetic valve
8
valve market
8
valve
7
34-year durability
4
debakey
4

Similar Publications

We report a case of a 57-year-old patient with a history of a Starr-Edwards mitral valve prosthesis and DeBakey-Surgitool aortic valve prosthesis implanted 30 years ago who presented with symptoms consistent with class IV heart failure. The patient had been on no anticoagulation for approximately 30 years secondary to recurrent epistaxis occurring two years after starting warfarin therapy postoperatively. Throughout the patient's lifetime he experienced no thromboembolic complications from the lack of anticoagulation, despite developing concomitant atrial fibrillation approximately ten years prior to admission.

View Article and Find Full Text PDF

34-year durability of a DeBakey Surgitool mechanical aortic valve prosthesis.

Interact Cardiovasc Thorac Surg

October 2006

Department of Cardiac Surgery, University Hospital Grosshadern, Ludwig-Maximilians-University, Marchioninistrasse 15, 81377 Munich, Germany.

Background: The DeBakey Surgitool mechanical aortic valve prosthesis was the first mechanical prosthetic valve in which pyrolitic carbon was used. This valve was on the market from 1969 to 1978 and approximately 3300 valves were implanted in this lapse worldwide.

Report: We present a DeBakey Surgitool mechanical aortic valve prosthesis implanted in 1972, which has been functioning for almost 34 years and still presents a good performance.

View Article and Find Full Text PDF

DeBakey Surgitool mechanical heart valve prosthesis, explanted at 32 years.

Cardiovasc Pathol

March 2005

Department of Pathology, University Health Network, Toronto General Hospital, 200 Elizabeth Street E4-301, Toronto, ON, Canada.

Contemporary mechanical heart value prostheses are expected to last "just about forever" or the patient's lifeline. They do however still suffer complications, some of which necessitate premature explantation. Complications today are mainly related to patient compliance with anticoagulant medication, infection and hemorrhage.

View Article and Find Full Text PDF

From August 1971 through November 1972, we implanted 62 Model 2 DeBakey-Surgitool aortic valve prostheses in 62 patients, 4 of whom later had clinically asymptomatic strut fractures. In 1 case, the patient died suddenly, and autopsy revealed detachment of the ball-cage; in each of the other 3 cases, fractures of 2 struts close to the base of the prosthesis were diagnosed fluoroscopically, and the patients underwent successful reoperation. The interval between implantation and reoperation ranged from 11 months to 16 years, 9 months.

View Article and Find Full Text PDF

Excessive cage strut wear allowing ball-poppet embolization caused the sudden death of a 47 year old lady in whom a DeBakey-Surgitool aortic prosthesis had been implanted nine years earlier. Patients with this type of prosthesis should have periodic valvular cine fluoroscopy with image intensification to allow visualization of significant strut wear or fracture, and appropriate prosthetic valve replacement.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!