Optical extinction monitor using cw cavity enhanced detection.

Rev Sci Instrum

Center for Sensor Systems and Technology, Aerodyne Research, Inc., 45 Manning Road, Billerica, MA 01821-3976, USA.

Published: June 2007

We present details of an apparatus capable of measuring optical extinction (i.e., scattering and/or absorption) with high precision and sensitivity. The apparatus employs one variant of cavity enhanced detection, specifically cavity attenuated phase shift spectroscopy, using a near-confocal arrangement of two high reflectivity (R approximately 0.9999) mirrors in tandem with an enclosed cell 26 cm in length, a light emitting diode (LED), and a vacuum photodiode detector. The square wave modulated light from the LED passes through the absorption cell and is detected as a distorted wave form which is characterized by a phase shift with respect to the initial modulation. The amount of that phase shift is a function of fixed instrument properties-cell length, mirror reflectivity, and modulation frequency-and of the presence of a scatterer or absorber (air, particles, trace gases, etc.) within the cell. The specific implementation reported here employs a blue LED; the wavelength and spectral bandpass of the measurement are defined by the use of an interference filter centered at 440 nm with a 20 nm wide bandpass. The monitor is enclosed within a standard 19 in. rack-mounted instrumentation box, weighs 10 kg, and uses 70 W of electrical power including a vacuum pump. Measurements of the phase shift induced by Rayleigh scattering from several gases (which range in extinction coefficient from 0.4-32 Mm(-1)) exhibit a highly linear dependence (r(2)=0.999 97) when plotted as the co-tangent of the phase shift versus the expected extinction. Using heterodyne demodulation techniques, we demonstrate a detection limit of 0.04 Mm(-1) (4 x 10(-10) cm(-1)) (2sigma) in 10 s integration time and a base line drift of less than +/-0.1 Mm(-1) over a 24 h period. Detection limits decrease as the square root of integration time out to approximately 150 s.

Download full-text PDF

Source
http://dx.doi.org/10.1063/1.2744223DOI Listing

Publication Analysis

Top Keywords

phase shift
20
optical extinction
8
cavity enhanced
8
enhanced detection
8
integration time
8
phase
5
shift
5
extinction monitor
4
monitor cavity
4
detection
4

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!