Background And Purpose: In helical tomotherapy the nature of the optimizing and planning systems allows the delivery of dose on the skin using a build-up compensating technique (skin flash). However, positioning errors or changes in the patient's contour can influence the correct dosage in these regions. This work studies the behavior of skin-flash regions using phantom and in-vivo dosimetry.

Materials And Methods: The dosimetric accuracy of the tomotherapy planning system in skin-flash regions is checked using film and TLD on phantom. Positioning errors are induced and the effect on the skin dose is investigated. Further a volume decrease is simulated using bolus material and the results are compared.

Results: Results show that the tomotherapy planning system calculates dose on skin regions within 2 SD using TLD measurements. Film measurements show drops of dose of 2.8% and 26% for, respectively, a 5mm and 10mm mispositioning of the phantom towards air and a dose increase of 9% for a 5mm shift towards tissue. These measurements are confirmed by TLD measurements. A simulated volume reduction shows a similar behavior with a 2.6% and 19.4% drop in dose, measured with TLDs.

Conclusion: The tomotherapy system allows adequate planning and delivery of dose using skin flashes. However, exact positioning is crucial to deliver the dose at the exact location.

Download full-text PDF

Source
http://dx.doi.org/10.1016/j.radonc.2007.06.003DOI Listing

Publication Analysis

Top Keywords

dose skin
12
skin flashes
8
helical tomotherapy
8
phantom in-vivo
8
dose
8
delivery dose
8
positioning errors
8
skin-flash regions
8
tomotherapy planning
8
planning system
8

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!