Objective: Optimization of cardiac resynchronization therapy (CRT) is still unsolved. It has been shown that optimal electrode position,atrioventricular (AV) and interventricular (VV) delays improve the success of CRT and reduce the number of non-responders. However, no automatic, noninvasive optimization strategy exists to date.
Methods: Cardiac resynchronization therapy was simulated on the Visible Man and a patient data-set including fiber orientation and ventricular heterogeneity. A cellular automaton was used for fast computation of ventricular excitation. An AV block and a left bundle branch block were simulated with 100%, 80% and 60% interventricular conduction velocity. A right apical and 12 left ventricular lead positions were set. Sequential optimization and optimization with the downhill simplex algorithm (DSA) were carried out. The minimal error between isochrones of the physiologic excitation and the therapy was computed automatically and leads to an optimal lead position and timing.
Results: Up to 1512 simulations were carried out per pathology per patient. One simulation took 4 minutes on an Apple Macintosh 2 GHz PowerPC G5. For each electrode pair an optimal pacemaker delay was found. The DSA reduced the number of simulations by an order of magnitude and the AV-delay and VV - delay were determined with a much higher resolution. The findings are well comparable with clinical studies.
Conclusion: The presented computer model of CRT automatically evaluates an optimal lead position and AV-delay and VV-delay, which can be used to noninvasively plan an optimal therapy for an individual patient. The application of the DSA reduces the simulation time so that the strategy is suitable for pre-operative planning in clinical routine. Future work will focus on clinical evaluation of the computer models and integration of patient data for individualized therapy planning and optimization.
Download full-text PDF |
Source |
---|
Heart Rhythm
January 2025
Geisinger Heart Institute, Geisinger Wyoming Valley Medical Center, MC 36-10, 1000 E Mountain Blvd, Wilkes-Barre, PA 18711.
Micromachines (Basel)
January 2025
Section of Electrophysiology, Division of Cardiology, Department of Internal Medicine, Rush University Medical Center, 1653 W. Congress, Chicago, IL 60612, USA.
Although significant strides have been made in cardiac pacing, the field is still evolving. While transvenous permanent pacing is highly effective in the management of bradyarrhythmias, it is not risk free and may result in significant morbidity and, rarely, mortality. Transvenous leads are often the weakest link in a pacing system.
View Article and Find Full Text PDFJ Cardiovasc Electrophysiol
January 2025
Department of Cardiac Electrophysiology and Pacing, Arrhythmia Heart Failure Academy, The Madras Medical Mission, Chennai, Tamil Nadu, India.
Introduction: Permanent implantation of a DF-4 implantable cardiac defibrillator (ICD) lead in the left bundle branch area (LBBA-ICD) is the next paradigm in amalgamating cardiac resynchronization therapy (CRT) and defibrillation. We systematically investigated feasibility/success rate, procedural caveats, and complications associated with a permanent DF-4 LBBA ICD implant and pertinent data at short-term follow-up.
Methods: We prospectively attempted implantation of 7 Fr Durata (Abbott, Chicago, IL, USA) single coil DF-4 ICD lead at the LBBA using a fixed-curve non-deflectable CPS locator delivery sheath.
Nucl Med Commun
January 2025
Department of Cardiac and Vascular Diseases, St.John Paul II Hospital, Jagiellonian University Medical College, Krakow, Poland.
Objectives: Cardiac resynchronization therapy (CRT) is an intervention for heart failure patients with reduced ejection fraction who exhibit specific electrocardiographic indicators of electrical dyssynchrony. However, electrical dyssynchrony does not universally correspond to left ventricular mechanical dyssynchrony (LVMD). Gated single-photon emission computed tomography (SPECT) myocardial perfusion allows for the assessment of LVMD, yet its role in the CRT selection process remains debated.
View Article and Find Full Text PDFFront Cardiovasc Med
January 2025
Department of Cardiology, The First People's Hospital of Neijiang, Neijiang, China.
More than 1 million permanent pacemakers are implanted worldwide each year, half of which are in patients with high-grade atrioventricular block. Pacemakers provide adequate frequency support in the initial stage, but traditional right ventricular (RV) pacing may lead to or aggravate left ventricular dysfunction and arrhythmia. Several potential risk factors for heart failure and arrhythmias after pacemaker surgery have been identified, but their occurrence remains difficult to predict clinically.
View Article and Find Full Text PDFEnter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!