Objective: Electrical impedance tomography (EIT) involves the application of a small alternating current to produce a series of chest images that can be used to monitor breathing pattern. The relation between chest images and tidal volume has not been sufficiently validated. The aim of the present study was to analyze the correlation between EIT images and the volume-time signal measured with a pneumotachometer in 13 healthy volunteers.
Material And Methods: The following measurement devices were used: a) MedGraphics preVent Pneumotach, with special software for recording the volume-time signal (reference test), and b) EIT-4, a fourth-generation prototype unit designed by the Department of Electronic Engineering at the Universidad Politécnica de Cataluña, Spain that records the volume-time signal and produces a graphic depiction of a cross section of the thorax at the sixth intercostal space.
Results: The mean (SD) tidal volume measured by the pneumotachometer and the EIT-4 was 0.523 (0.102) L and 0.527 (0.106) L, respectively (P value not significant). The linear correlation coefficient between the 2 measurements was 0.923 (P=.001), and the mean of the differences between the 2 procedures was -0.003 L (95% confidence interval, -0.045 to 0.038). The greatest differences were associated with female gender, body mass index, and chest circumference. In view of these differences, a different equation based on these variables was needed for calibration of the EIT-4.
Conclusions: The EIT-4 provides an alternative means of monitoring breathing pattern, although a number of issues related to the circumference of the rib cage need to be resolved.
Download full-text PDF |
Source |
---|---|
http://dx.doi.org/10.1016/s1579-2129(07)60074-3 | DOI Listing |
Sci Rep
January 2025
Department of Mechanical Engineering, PSG Institute of Technology and Applied Research, Coimbatore, 641026, India.
Typical waveforms used for the simulation of pressure and volume-controlled ventilation in medical ventilators have been extensively studied in the literature. The majority of simulation studies reported employ the step pattern or ramp pattern to model the pressure and flow variations in pressure/volume-controlled ventilation. It was observed that the above waveforms tend to add to the discomfort level of patients due to the presence of jerks in derivatives of pressure/flow variations; the pressure/flow variation of air and oxygen mixture should be smooth so that the patient discomfort is kept at a minimal level.
View Article and Find Full Text PDFRev Col Bras Cir
January 2025
- School of Medical Sciences Orebro university, Department of Surgery - Orebro - OR - Suécia.
Introduction: Hemorrhage is the leading cause of preventable deaths in trauma patients, resulting in 1.5 million deaths annually worldwide. Traditional trauma assessment follows the ABC (airway, breathing, circulation) sequence; evidence suggests the CAB (circulation, airway, breathing) approach to maintain perfusion and prevent hypotension.
View Article and Find Full Text PDFCrit Care Explor
January 2025
All authors: Department of Pharmacy, Brigham and Women's Hospital, Boston, MA.
Importance: Recent studies have found an association between COVID-19 infection and deeper sedation in mechanically ventilated patients, raising concerns about the impact of the COVID-19 pandemic on pain, agitation, and delirium (PAD) management practices overall.
Objectives: This study aimed to assess differences in PAD management in patients without COVID-19 infection in pre- and peri-COVID-19 pandemic timeframes.
Design, Setting, And Participants: This was a single-center, retrospective, pre-/post-cohort analysis of mechanically ventilated adult patients without COVID-19 infection admitted to an ICU in Boston, MA.
Brain Struct Funct
January 2025
Department of Psychiatry, Psychotherapy and Psychosomatics, School of Medicine, RWTH Aachen University, Aachen, Germany.
Physiological responses derived from audiovisual perception during assisted driving are associated with the regulation of the autonomic nervous system (ANS), especially in emergencies. However, the interaction of event-related brain activity and the ANS regulating peripheral physiological indicators (i.e.
View Article and Find Full Text PDFAging Dis
January 2025
Department of Medical Oncology, Shanghai Pulmonary Hospital, School of Medicine, Tongji University, Shanghai, China.
Circadian rhythm is the internal homeostatic physiological clock that regulates the 24-hour sleep/wake cycle. This biological clock helps to adapt to environmental changes such as light, dark, temperature, and behaviors. Aging, on the other hand, is a process of physiological changes that results in a progressive decline in cells, tissues, and other vital systems of the body.
View Article and Find Full Text PDFEnter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!