This work aims at presenting a study using Monte Carlo simulation of a Multileaf Shielding (MLS) System designed to be used for the protection of patients who undergo radiotherapy treatment, against undesired exposure to neutrons produced in the components of the medical linear accelerator heads. The choice of radiotherapy equipment as the subject of study fell on the Varian Clinac 2,100/2,300 with MLC-120 operating at 18 MeV. The general purpose Monte Carlo N-Particle radiation transport code, MCNP5, was used in the computer simulation in order to determine the ambient dose equivalent, H (10), on several points on the patient's plane, with the equipment operation with and without the MLS. The results of the simulations showed a significant neutron dose reduction after the inclusion of the proposed shielding.
Download full-text PDF |
Source |
---|---|
http://dx.doi.org/10.1093/rpd/ncm312 | DOI Listing |
J Appl Clin Med Phys
November 2024
Department of Radiation Oncology, Shanghai Concord Medical Cancer center, Shanghai, China.
J Appl Clin Med Phys
September 2024
Département de Radio-oncologie, CIUSSS de l'Est-de-L'Île-de-Montréal, Hôpital Maisonneuve-Rosemont, Montréal, Quebec, Canada.
Purpose: The purpose of this work was to detail our center's experience in transitioning from a Co-60 treatment technique to an intensity modulated radiation therapy (IMRT) based lateral-field extended source-to-axis distance (e-SAD) technique for total body irradiation (TBI).
Materials And Methods: An existing beam model in RayStation v.10A was validated for the use of e-SAD TBI treatments.
J Appl Clin Med Phys
March 2024
Department of Radiation Oncology, School of Medicine, Stanford University, Stanford, California, USA.
Purpose: A well-known limitation of multi-leaf collimators is that they cannot easily form island blocks. This can be important in mantle region therapy. Cerrobend photon blocks, currently used for supplementary shielding, are labor-intensive and error-prone.
View Article and Find Full Text PDFJ Cancer Res Ther
November 2023
Seyed-Al-Shohada Hospital, Isfahan University of Medical Science; Department of Radio-Oncology, Seyed-Al-Shohada Hospital, Isfahan University of Medical Sciences, Isfahan, Iran.
Introduction: Radiotherapy (RT) may increase the dose of heart structure like left anterior descending coronary artery (LAD). The purpose of this paper was to evaluate the impact of various multileaf collimators (MLCs) in shielding organ at risks (OARs), especially LAD, of patients with left breast cancer.
Materials And Methods: Forty-five patients with left breast cancer were selected.
Adv Biomed Res
April 2023
Department of Radio-Oncology, Seyed-Al-Shohada Hospital, Isfahan University of Medical Sciences, Isfahan, Iran.
Background: Multi-leaf collimator (MLC) is one of the efficient and cost-effective methods for protecting sensitive tissues around the target. This study aimed to evaluate the protective effect of MLC on the protection of sensitive organs in patients with left breast cancer.
Materials And Methods: This study was performed on computed tomography (CT) scans of 45 patients with left breast cancer.