We evaluated the performance characteristics of three real-time reverse transcription-PCR test systems for detection and quantification of hepatitis C virus (HCV) and performed a direct comparison of the systems on the same clinical specimens. Commercial HCV panels (genotype 1b) were used to evaluate linear range, sensitivity, and precision. The Roche COBAS TaqMan HCV test for research use only (RUO) with samples processed on the MagNA Pure LC instrument (Roche RUO-MPLC) and Abbott analyte-specific reagents (ASR) with QIAGEN sample processing (Abbott ASR-Q) showed a sensitivity of 1.0 log(10) IU/ml with a linear dynamic range of 1.0 to 7.0 log(10) IU/ml. The Roche ASR in combination with the High Pure system (Roche ASR-HP) showed a sensitivity of 1.4 log(10) IU/ml with a linear dynamic range of 2.0 to 7.0 log(10) IU/ml. All of the systems showed acceptable reproducibility, the Abbott ASR-Q being the most reproducible of the three systems. Seventy-six clinical specimens (50 with detectable levels of HCV RNA and various titers and genotypes) were tested, and results were compared to those of the COBAS Amplicor HCV Monitor v2.0. Good correlation was obtained for the Roche RUO-MPLC and Abbott ASR-Q (R(2) = 0.84 and R(2) = 0.93, respectively), with better agreement for the Abbott ASR-Q. However, correlation (R(2) = 0.79) and agreement were poor for Roche ASR-HP, with bias relative to concentration and genotype. Roche ASR-HP underestimated HCV RNA for genotypes 3 and 4 as much as 2.19 log(10) IU/ml. Our study demonstrates that Roche RUO-MPLC and Abbott ASR-Q provided acceptable results and agreed sufficiently with the COBAS Amplicor HCV Monitor v2.0.

Download full-text PDF

Source
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC1951251PMC
http://dx.doi.org/10.1128/JCM.00058-07DOI Listing

Publication Analysis

Top Keywords

abbott asr-q
20
log10 iu/ml
20
roche ruo-mplc
12
ruo-mplc abbott
12
roche asr-hp
12
performance characteristics
8
characteristics three
8
three real-time
8
real-time reverse
8
reverse transcription-pcr
8

Similar Publications

We evaluated the performance characteristics of three real-time reverse transcription-PCR test systems for detection and quantification of hepatitis C virus (HCV) and performed a direct comparison of the systems on the same clinical specimens. Commercial HCV panels (genotype 1b) were used to evaluate linear range, sensitivity, and precision. The Roche COBAS TaqMan HCV test for research use only (RUO) with samples processed on the MagNA Pure LC instrument (Roche RUO-MPLC) and Abbott analyte-specific reagents (ASR) with QIAGEN sample processing (Abbott ASR-Q) showed a sensitivity of 1.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!