Introduction: The use of magnetic fields to treat disease has intrigued mankind since the time of the ancient Greeks. More recently it has been shown that electromagnetic field (EMF) treatment aids bone healing, and repetitive transcranial magnetic stimulation (rTMS) appears to be beneficial in treating schizophrenia and depression. Since external EMFs influence internal body processes, we hypothesized that measurement of body surface EMFs might be used to detect disease states and direct the course of subsequent therapy. However, measurement of minute body surface EMFs requires use of a sensitive and well documented magnetometer. In this study we evaluated the sensitivity and frequency response of a fluxgate magnetometer with a triaxial probe for use in detecting body surface EMF and we compared the magnetometer readings with a signal from a Toftness Sensometer, operated by an experienced clinician, in the laboratory and in a clinical setting.

Methods: A Peavy Audio Amplifier and variable power output Telulex signal generator were used to develop 50 microT EMFs in a three coil Merritt coil system. A calibrated magnetometer was used to set a 60 Hz 50 microT field in the coil and an ammeter was used to measure the current required to develop the 50 microT field. At frequencies other than 60 Hz, the field strength was maintained at 50 microT by adjusting the Telulex signal output to keep the current constant. The field generated was monitored using a 10 turn coil connected to an oscilloscope. The oscilloscope reading indicated that the field strength was the same at all frequencies tested. To determine if there was a correspondence between the signals detected by a fluxgate magnetometer (FGM1) and the Toftness Sensometer both devices were placed in the Merritt coil and readings were recorded from the FGM1 and compared with the ability of a highly experienced Toftness operator to detect the 50 microT field. Subsequently, in a clinical setting, FGM1 readings made by an FGM1 technician and Sensometer readings were made by 4 Toftness Sensometer operators, having various degrees of experience with this device. Each examiner obtained instrument readings from 5 different volunteers in separate chiropractic adjusting rooms. Additionally, one of the Toftness Sensometers was equipped with an integrated fluxgate magnetometer (FGM2) and this magnetometer was used to obtain a second set of EMF readings in the clinical setting.

Results: The triaxial fluxgate magnetometer was determined to be moderately responsive to changes in magnetic field frequency below 10 Hz. At frequencies above 10 Hz the readings corresponded to that of the ambient static geofield. The practitioner operating the Toftness Sensometer was unable to detect magnetic fields at high frequencies (above 10 Hz) even at very high EMFs. The fluxgate magnetometer was shown to be essentially a DC/static magnetic field detector and like all such devices it has a limited frequency range with some low level of sensitivity at very low field frequencies. The interexaminer reliability of four Toftness practitioners using the Sensometer on 5 patients showed low to moderate correlation.

Conclusions: The fluxgate magnetometer although highly sensitive to static (DC) EMFs has only limited sensitivity to EMFs in the range of 1 to 10 Hz and is very insensitive to frequencies above 10 Hz. In laboratory comparisons of the Sensometer and the fluxgate magnetometer there was an occasional correspondence between the two instruments in detecting magnetic fields within the Merritt coil but these occasions were not reproducible. In the clinical studies there was low to moderate agreement between the clinicians using the Sensometer to diagnosing spinal conditions and there was little if any agreement between the Sensometer and the fluxgate magnetometer in detecting EMFs emanating from the volunteers body surface.

Download full-text PDF

Source
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC1769295PMC

Publication Analysis

Top Keywords

fluxgate magnetometer
36
toftness sensometer
20
body surface
20
magnetometer
13
magnetic fields
12
merritt coil
12
microt field
12
sensometer
10
field
10
fluxgate
9

Similar Publications

Applicability of Small and Low-Cost Magnetic Sensors to Geophysical Exploration.

Sensors (Basel)

October 2024

Department of Earth, Environmental and Resources Sciences, University of Naples "Federico II", 80126 Naples, Italy.

In the past few decades, there has been a notable technological advancement in geophysical sensors. In the case of magnetometry, several sensors were used, having the common feature of being miniaturized and lightweight, thus idoneous to be carried by UAVs in drone-borne magnetometric surveys. A common feature is that their sensitivity ranges from 0.

View Article and Find Full Text PDF

Structural Design and Parameter Optimization of Magnetic Gradient Tensor Measurement System.

Sensors (Basel)

June 2024

State Key Laboratory of Dynamic Measurement Technology, North University of China, Taiyuan 030051, China.

Magnetic anomaly detection (MAD) technology based on the magnetic gradient tensor (MGT) has broad application prospects in fields such as unexploded ordnance detection and mineral exploration. The difference approximation method currently employed in the MGT measurement system introduces measurement errors. Designing reasonable geometric structures and configuring optimal structural parameters can effectively reduce measurement errors.

View Article and Find Full Text PDF

The highest level of the present-day geomagnetic recordings is presented by the absolute controlled magnetic measurements. This quality is permanently fulfilled in geomagnetic observatories (GO) only. The absolute records are based on nuclear magnetometers for the intensity measurement and on DIM (Declination and Inclination Magnetometer) instruments for the direction determination of the ambient magnetic field.

View Article and Find Full Text PDF

This research proposes a magnetic field sensor with spatial orientation ability. Through the assistance of a magnetic flux concentrator, out-of-plane magnetic flux can be concentrated and guided into the planar magnetic cores of a fluxgate sensor. A printed circuit board is used to construct the basic planar structure, on which the proposed three-dimensional magnetic flux concentrator and magnetic cores are assembled.

View Article and Find Full Text PDF

Triaxial magnetic field coils are one of the most important components of magnetic resonance sensors. Traditional measurement methods for coil constants and non-orthogonal angles using fluxgate magnetometers are no longer suitable for small-volume nuclear magnetic resonance sensors. This study presents a method for measuring the coil constants and nonorthogonal angles of triaxial nuclear magnetic resonance coils using the dynamics of the electron paramagnetic resonance magnetometer without requiring any additional calibration equipment.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!