Complication risk with pulse generator change: implications when reacting to a device advisory or recall.

Pacing Clin Electrophysiol

Division of Cardiology, Mayo Clinic, Rochester, Minnesota 55905, USA.

Published: June 2007

Background: Recent advisories and recalls of pacemakers and implantable cardioverter-defibrillators (ICDs) have highlighted the need for evidence-based recommendations regarding management of patients with advisory devices. In order to better facilitate decision-making when weighing the relative risks and benefits of performing generator changes in these patients, we conducted a review to assess operative complication rates.

Methods: We reviewed generator changes performed between 2000 and 2005 at the Mayo Clinic-Rochester, including a total of 732 change-outs consisting of 570 done for elective replacement indicators (ERI) and 162 for manufacturer advisories or recalls. Complications included all those requiring reoperation, occurring within a 60-day period postoperatively and directly attributable to the generator change. These included infection requiring device excision, hematoma requiring evacuation, and incisional dehiscence requiring reclosure.

Results: Operation-associated complications requiring intervention were noted in 9 patients, or 1.24% of our population. Of these nine complications, eight occurred among patients receiving pulse generator replacement for ERI (1.40%) and one occurred in a patient receiving replacement for a manufacturer advisory or recall (0.62%). Complications included 5 infections, 3 hematomas, and 1 incisional dehiscence.

Conclusions: Generator replacement is not a benign procedure and associated risks must be weighed in the context of other variables when making management choices in patients with advisory or recall devices.

Download full-text PDF

Source
http://dx.doi.org/10.1111/j.1540-8159.2007.00742.xDOI Listing

Publication Analysis

Top Keywords

advisory recall
12
pulse generator
8
generator change
8
advisories recalls
8
patients advisory
8
generator changes
8
complications included
8
generator replacement
8
generator
6
patients
5

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!