Acoustic droplet vaporization threshold: effects of pulse duration and contrast agent.

IEEE Trans Ultrason Ferroelectr Freq Control

University of Michigan, Department of Biomedical Engineering and Radiology, Ann Arbor, MI, USA.

Published: May 2007

The use of superheated liquid perfluorocarbon droplets encased in albumin shells has been proposed as a minimally invasive alternative to current treatment of cancer by means of occlusion therapy. In response to an applied acoustic field, these droplets, which are small enough to pass through capillaries, vaporize into large gas bubbles that subsequently lodge in the vasculature. This technique, known as acoustic droplet vaporization (ADV) has been shown to successfully reduce blood flow in vivo, but for in situ conditions where attenuation is present, lower acoustic frequency and ADV threshold may be desirable. Thus, two methods to lower the ADV threshold at a lower 1.44 MHz were explored. The first part of this study investigated the role of pulse duration on ADV. The second part investigated the role of inertial cavitation (IC) external to a droplet by lowering the IC threshold in the host liquid with the presence of Definity contrast agent (CA). The threshold was found to be 5.5-5.9 MPa for short microsecond pulses and decreased for millisecond pulses (3.8-4.6 MPa). When CAs were present and long millisecond pulses were used, the ADV threshold decreased to values as low as 0.41 MPa.

Download full-text PDF

Source
http://dx.doi.org/10.1109/tuffc.2007.339DOI Listing

Publication Analysis

Top Keywords

adv threshold
12
acoustic droplet
8
droplet vaporization
8
pulse duration
8
contrast agent
8
investigated role
8
millisecond pulses
8
threshold
6
adv
5
acoustic
4

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!