Linearity of the dose monitor system at low monitor units.

Strahlenther Onkol

Strahlentherapie, Diakoniekrankenhaus Schwäbisch Hall, Diakoniestrasse 10, 74523 Schwäbisch Hall, Germany.

Published: June 2007

Background And Purpose: The accuracy of the dose monitor system of a medical electron accelerator is of crucial importance for the precise delivery of the prescribed dose. Special attention is given to the case of low monitor units (MU). The long-term stability of the dose monitor system is analyzed.

Material And Methods: Photon and electron fields with monitor units between 1 and 500 MU were measured with field sizes of 10 x 10 cm(2) and 20 x 20 cm(2) at source-skin distances (SSDs) of 100 cm and 95 cm, respectively. The relative dose per MU [see text] was determined for an Elekta linear accelerator.

Results: The relative dose per MU was measured to be constant within about 1% down to fields with only 2 MU. The deviation remained below 2% down to 1 MU which is the minimum setting of the accelerator. The long-term stability is better than 0.2% per month.

Conclusion: No therapeutically relevant deviation of the prescribed dose could be detected for the Elekta linear accelerator. This is a prerequisite for the implementation of intensity-modulated radiotherapy (IMRT) techniques.

Download full-text PDF

Source
http://dx.doi.org/10.1007/s00066-007-1596-2DOI Listing

Publication Analysis

Top Keywords

dose monitor
12
monitor system
12
monitor units
12
low monitor
8
prescribed dose
8
long-term stability
8
relative dose
8
elekta linear
8
monitor
6
dose
6

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!