Purpose: To determine the interreader and intrareader agreement at the Optical Coherence Tomography (OCT) Reading Center at Duke for images produced for an interventional neovascular age-related macular degeneration (AMD) clinical trial.

Design: Retrospective, observational case series.

Methods: OCT was performed using the Stratus OCT Fast Macular Thickness Map (Carl Zeiss Meditec, Dublin, California, USA) scan mode and a 7-mm line scan centered on the fovea. Experienced OCT readers, certified to grade scans according to a standardized protocol, independently determined whether a scan was gradable. If gradable, each of the scans was graded for multiple morphologic characteristics. In addition, retinal thickness, subretinal fluid thickness, and choroidal neovascularization (CNV) thickness at the fovea were measured for each gradable scan. Interreader agreement was determined among three reader pairs. Readers regraded a subset of scans and intrareader agreement was determined.

Results: The interreader agreement was high for scan gradability and scan grades among three reader pairs, ranging from 84% to 100% and from 84% to 96%, respectively. Similarly, the intrareader agreement for scan gradability and scan grade, including comparison of adjudicated scan grades with readjudicated scan grades, was high and ranged from 91% to 100% and from 79% to 98%, respectively, except for scan grades of retinal pigment epithelium atrophy, when CNV was present. Highly reproducible results also were found for quantitative thickness measurements.

Conclusions: Well-trained OCT readers can grade independently, with a high level of interreader and intrareader agreement, multiple morphologic parameters of OCT scans obtained from eyes with neovascular AMD. Reproducible Stratus OCT scan data will be valuable to monitor treatment efficacy in interventional clinical trials of neovascular AMD.

Download full-text PDF

Source
http://dx.doi.org/10.1016/j.ajo.2007.03.056DOI Listing

Publication Analysis

Top Keywords

intrareader agreement
16
scan grades
16
scan
12
optical coherence
8
coherence tomography
8
neovascular age-related
8
age-related macular
8
macular degeneration
8
interreader intrareader
8
stratus oct
8

Similar Publications

Background: Image-defined risk factors (IDRFs) were introduced to provide a consensus approach for pre-treatment risk stratification on computed tomography (CT) and magnetic resonance imaging (MRI) in patients with neuroblastoma.

Objective: To assess the intra- and inter-reader agreement of radiologists in identifying IDRFs on CT.

Materials And Methods: Approval for this retrospective study was granted by our institutional research ethics board with a waiver of consent.

View Article and Find Full Text PDF

Ultra-fast [F]florbetapir PET imaging using the uMI Panorama PET/CT system.

EJNMMI Phys

December 2024

Department of Nuclear Medicine, State Key Laboratory of Complex Severe and Rare Diseases, Center for Rare Diseases Research, Beijing Key Laboratory of Molecular Targeted Diagnosis and Therapy in Nuclear Medicine, Peking Union Medical College Hospital, Chinese Academy of Medical Science and Peking Union Medical College, 1# Shuaifuyuan, Dongcheng District, Beijing, 100730, China.

Background: There is a need for faster amyloid PET scans to reduce patients' discomfort, minimize movement artifacts, and increase throughput. The recently introduced uMI Panorama PET/CT system featuring enhanced spatial resolution and sub-200ps TOF offers the potential for shorter scan duration without sacrificing image quality or efficacy to detect Aβ deposition. The study aims to establish a faster acquisition protocol for [F]florbetapir PET imaging using digital PET/CT scanner uMI Panorama, while ensuring adequate image quality and amyloid-β (Aβ) detectability comparable to the standard 10-minute scan.

View Article and Find Full Text PDF

A four-dimensional phase-contrast magnetic resonance imaging sequence with respiratory-controlled adaptive k-space reordering (ReCAR-4DPC) offers potential benefits of improved scan efficiency and motion robustness. The purpose of this study was to evaluate the reproducibility of flow measurement using this technique and to compare hemodynamic metrics obtained to two-dimensional phase contrast MRI (2DPC)-derived metrics of the thoracic aorta. ReCAR-4DPC was performed with identical scan parameters in 15 healthy volunteers (6M,9F, mean [range] 37 [23-47] years) and 11 patients with thoracic aortic dissection (6M,5F, 56 [31-81] years) and acquisition time was recorded.

View Article and Find Full Text PDF

Objective: To compare abbreviated magnetic resonance imaging (MRI) to complete MRI for treatment response assessment of colorectal liver metastases.

Material And Methods: This retrospective study included consecutive patients with colorectal liver metastases between January 1, 2012, and December 3, 2021, who were undergoing chemotherapy and who had at least one follow-up gadoxetic-enhanced MRI. For each patient, two MRIs (baseline MRI and follow-up MRI) were randomly selected.

View Article and Find Full Text PDF

Purpose: Breast density assessment is an essential component of risk-stratified and supplemental breast cancer screenings for the future generation. International collaborative research and cooperation are required to determine global trends and differences in mammographic breast density. Due to the cost and energy associated with the use of film or digital mammograms, international breast density assessment studies tend to be challenging.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!