Objective: Various types of rotary blood pumps (axial flow, centrifugal) have been introduced into clinical use recently. These pumps have different pressure-flow characteristics, and some investigators have noted that a limited pump flow rate and less pulsatility are the problems with the axial flow devices.

Methods: A new implantable centrifugal blood pump was developed that has an extremely flat pressure-flow curve and is able to produce a significantly high pump flow rate of 20 l/min at a low pressure of 10-30mmHg. When the pressure difference between the left ventricle and aorta decreases during systole, an instant high peak flow is achieved, which results in a higher peak pressure in the aorta (systolic pressure). During the diastolic phase, the left ventricle-aorta pressure difference increases to maximum, and the pump flow rate decreases to minimum. Thus, the pump flow rate becomes completely pulsatile, and the high peak flow provides a higher mean pump flow rate. This pump was applied to two end-stage heart failure patients (dilated cardiomyopathy, New York Heart Association (NYHA) class IV, inotrope-dependent).

Results: The pump was observed to provide completely pulsatile high flow assistance of 6-9 l/min with a constant pump speed. Both patients are currently in NYHA class I after 1 year on the device with no major adverse events.

Conclusion: The new centrifugal blood pump provides completely pulsatile high-flow circulatory support with a constant pump speed, which solves the current clinical problems with rotary blood pumps.

Download full-text PDF

Source
http://dx.doi.org/10.1007/s11748-007-0098-6DOI Listing

Publication Analysis

Top Keywords

pump flow
20
flow rate
20
completely pulsatile
16
pulsatile high
12
centrifugal blood
12
pump
12
blood pump
12
flow
11
high flow
8
circulatory support
8

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!