Purpose: To compare intraocular pressure (IOP) measurements taken with Pascal dynamic contour tonometry (DCT), the TonoPen and the Goldmann applanation tonometry (GAT). The influence of central corneal thickness (CCT) on IOP measurements taken with Pascal DCT and the TonoPen was evaluated.
Methods: One eye in each of 101 consecutive patients with primary open-angle glaucoma (POAG) underwent ultrasonic CCT measurement and IOP evaluation with GAT, Pascal DCT and the TonoPen in random order. The agreement between results from Pascal DCT and the TonoPen and those of GAT was assessed using the Bland-Altman method. The deviation of Pascal DCT and TonoPen readings from GAT values, corrected for CCT, was calculated and correlated to CCT using a linear regression model.
Results: The mean of the differences in IOP measurements was 3.2 +/- 2.4 mmHg for Pascal DCT minus GAT readings and 0.5 +/- 4.5 mmHg for TonoPen minus GAT readings. The 95% confidence interval of differences in IOP measurements was higher between TonoPen and GAT readings (- 6 to 7 mmHg) than between Pascal and GAT readings (0.1-6.8 mmHg). Pascal DCT significantly overestimated IOP compared with GAT, especially for higher IOP readings. Bland-Altman scatterplots showed reasonable inter-method agreement between Pascal DCT and GAT measurements, and poor agreement between TonoPen and GAT measurements. The deviations of Pascal DCT and TonoPen readings from the corrected GAT values were both highly correlated with CCT values (linear regression analysis, p < 0.0001). The mean change in measured IOP for a 10-microm increase in CCT was 0.48 mmHg for Pascal DCT and 0.74 mmHg for the TonoPen.
Conclusions: Agreement with GAT measurements was higher for Pascal DCT than for TonoPen readings; however, Pascal DCT significantly overestimated IOP values compared with GAT. Measurements of IOP obtained with both Pascal DCT and the TonoPen appeared to be influenced by CCT, and this influence appeared to be greater for the latter.
Download full-text PDF |
Source |
---|---|
http://dx.doi.org/10.1111/j.1600-0420.2006.00834.x | DOI Listing |
To assess the impact of microgravity exposure on ocular rigidity (OR), intraocular pressure (IOP), and ocular pulse amplitude (OPA) following long-term space missions. OR was evaluated using optical coherence tomography (OCT) and deep learning-based choroid segmentation. IOP and OPA were measured with the PASCAL Dynamic Contour Tonometer (DCT).
View Article and Find Full Text PDFIndian J Ophthalmol
May 2024
Department of Ophthalmology, University Hospitals Leuven (UZ Leuven), Leuven, Belgium.
Purpose: A meta-analysis found that including atmospheric pressure as altitude in generalized linear models reveals higher differences between Goldmann tonometry and Pascal dynamic contour tonometry at higher altitudes, with the difference increasing in thinner corneas. To examine the difference in intraocular pressure (IOP) measurements by using Goldman applanation tonometry (GAT) and dynamic contour tonometer (DCT) tonometry in published literature and determine the influence of central corneal thickness (CCT), age, and altitude on that difference.
Methods: Articles that compare GAT and DCT were selected for an extensive literature review, and the location and altitude of the research centers were found online.
PLoS One
March 2023
Department of Biomedical Engineering, The Ohio State University, Columbus, OH, United States of America.
The purpose of this study was to develop a method of estimating pulsatile ocular blood volume (POBV) from measurements taken during an ophthalmic exam, including axial length and using a tonometer capable of measuring intraocular pressure (IOP) and ocular pulse amplitude (OPA). Unpublished OPA data from a previous invasive study was used in the derivation, along with central corneal thickness (CCT) and axial length (AL), as well as IOP from the PASCAL dynamic contour tonometer (DCT) and intracameral (ICM) measurements of IOP for 60 cataract patients. Intracameral mean pressure was set to 15, 20, and 35 mmHg (randomized sequence) in the supine position, using a fluid-filled manometer.
View Article and Find Full Text PDFInt Ophthalmol
March 2023
Department of Cardiology, Faculty of Medicine, Adnan Menderes University, Aydin, Turkey.
Purpose: To examine whether there is a relationship between ocular pulse amplitude (OPA), intraocular pressure (IOP), and visual field (VF) deterioration among the patients diagnosed with aortic regurgitation (AR).
Methods: Twenty-nine patients (average age of 62.06 ± 13.
J Glaucoma
November 2022
University of Health Sciences, Beyoglu Eye Training and Research Hospital, Istanbul, Turkey.
Prcis: IOP measured with Tonopen and DCT was higher than GAT in eyes that underwent DMEK.
Purpose/aim Of The Study: To compare intraocular pressure (IOP) measurements measured based on Goldmann applanation tonometry (GAT), Tonopen XL, and Pascal Dynamic Contour Tonometry (DCT) in patients who had undergone descemet membrane endothelial keratoplasty (DMEK) and to appraise the influence of central corneal thickness (CCT) on IOP measurements.
Materials And Methods: Thirty-four eyes (from 34 patients) who underwent DMEK at least 1 month before the study were included.