A disposable ion-selective optode for mercury based on trityl-picolinamide (T-Pico) as neutral ionophore was developed. The sensing layer consist of plasticised PVC incorporating T-Pico as a selective ionophore for Hg2+, ETH 5418 as a chromoionophore, and potassium tetrakis[3,5-bis(trifluoromethyl)phenyl] borate as lipophilic salt. The measurement principle is based on an ion-exchange mechanism. When the optode membrane is introduced into a water sample for 5 min, there is a colour change from red to blue, depending on the mercury concentration (pH 4.7), making it possible to use the absorbance at 665 nm as the analytical signal. The optode membrane response to Hg2+ was not fully reversible; however, it reveals a very good selectivity to many cations including alkali, alkaline earth, transition, and heavy metal ions. The detection limit for Hg2+ is 5.0x10(-7) M at pH 4.7. The response characteristics of the sensor including dynamic range, reproducibility, response time, and lifetime are discussed in detail. This sensor was used for the determination of mercury in environmental water samples with satisfactory recovery.

Download full-text PDF

Source
http://dx.doi.org/10.1016/j.aca.2007.03.064DOI Listing

Publication Analysis

Top Keywords

ion-selective optode
8
based trityl-picolinamide
8
optode membrane
8
development disposable
4
mercury
4
disposable mercury
4
mercury ion-selective
4
optode
4
optode based
4
trityl-picolinamide ionophore
4

Similar Publications

In recent years, ion-selective optodes (ISOs) have remarkably progressed, driven by innovative modern designs and nanomaterial integration. This review explored the development of modern ISO by describing state-of-the-art strategies to improve their sensitivity, selectivity, and real-time monitoring capacity. The review reported the traditional membrane based-optodes, and investigated the latest research, current design principles, and the use of essential components, such as ionophores, indicator dyes, polymer membranes, and nanomaterials, in ISO fabrication.

View Article and Find Full Text PDF

Exploiting the pH-Cross Sensitivity of Ion-Selective Optodes to Broaden Their Response Range.

ACS Sens

September 2024

Aarhus University Centre for Water Technology, Department of Biology, Section for Microbiology, Aarhus University, Ny Munkegade 114-1540, Aarhus C, 8000, Denmark.

While the pH cross-sensitivity of chromoionophore-based ion-selective optodes (ISOs) has often been regarded as a significant limitation, this paper demonstrates how this apparent drawback can be transformed into a beneficial feature. The response range of chromoionophore-based ISOs shifts proportionally with changes in the sample pH. Thus, integrating them with a stable pH gradient across the optode surface, such as those provided by immobilized pH gradient (IPG) gels, allows for significant enhancement of the effective measuring range of chromoionophore-based ISOs while preserving their maximum sensitivity.

View Article and Find Full Text PDF

Functionalized conducting polymers (FCPs) have recently garnered attention as ion-selective sensor materials, surpassing their intrinsic counterparts due to synergistic effects that lead to enhanced electrochemical and analytical parameters. Following a brief introduction of the fundamental concepts, this article provides a comprehensive review of the recent developments in the application of FCPs in ion-selective electrodes (ISEs) and ion-selective optodes (ISOs), particularly as ion-to-electron transducers, optical transducers, and ion-selective membranes. Utilizing FCPs in these devices offers a promising avenue for detecting and measuring ions in various applications, regardless of the sample nature and composition.

View Article and Find Full Text PDF

This study introduces a novel microfluidic polymer chip system that employs an embedded anionic surfactant (AS) ion-selective fluorescence optode (AS fluorescence optode) as a detector for measuring AS. The AS fluorescent optode comprises a lactone form of rhodamine B (L-RB) embedded in 2-nitrophenyl octyl ether plasticized poly (vinyl chloride) membrane. The AS fluorescence optode demonstrated a linear correlation between fluorescence intensity peak heights and AS concentrations within the range of less than 20 µM under optimal flow conditions.

View Article and Find Full Text PDF

Fluoride ions are highly relevant in environmental and biological sciences, and there is a very limited number of established fluoride chemical sensors. Previous fluoride-selective optodes were demonstrated with metal-porphyrin as the ionophore and required a chromoionophore for optical signal transduction. We demonstrate here novel optical fluoride sensing with nano-optodes containing an aluminum-phthalocyanine complex (AlClPc) as the single active sensing component, simplifying the conventional ion-selective optodes approach.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!