Objective: To compare experimentally the image quality of charged-coupled device (CCD)-based digital, flatbed scanner digitized, and dental film radiographs.

Study Design: High-contrast standardized radiographs of an aluminum step wedge containing boreholes were obtained on dental radiographic film (F) and a CCD receptor (D). Digitization (S) was done with a flatbed scanner. Each radiograph was separated into 23 single images, 18 of which showed a dark spot. Thirty observers indicated their confidence on the visibility of the spots for a receiver operating characteristic (ROC) analysis.

Results: Mean Az values (area beneath the ROC curves) differed significantly (P < .05), being highest for D (Az = 0.76), followed by F (Az = 0.71) and S (Az = 0.60). Mean sensitivity of S (0.45) was significantly (P = .000) lower than for D (0.66) and F (0.67).

Conclusion: The difference between modality D and F was small. However, the inferior performance of S images, particularly in dark regions, could be clinically relevant.

Download full-text PDF

Source
http://dx.doi.org/10.1016/j.tripleo.2006.04.005DOI Listing

Publication Analysis

Top Keywords

digitized dental
8
receiver operating
8
operating characteristic
8
flatbed scanner
8
images dark
8
vitro perception
4
perception low-contrast
4
low-contrast features
4
features digital
4
digital film
4

Similar Publications

Clinical characteristics and removal of broken burs retained in the lower jaw.

BMC Oral Health

January 2025

State Key Laboratory of Oral & Maxillofacial Reconstruction and Regeneration, Key Laboratory of Oral Biomedicine Ministry of Education, Hubei Key Laboratory of Stomatology, School & Hospital of Stomatology, Wuhan University, Wuhan, China.

Background: A broken bur retained in the lower jaw is an uncommon complication that occurs during the extraction of the impacted mandibular third molar. The purpose of this retrospective study was to investigate the clinical characteristics of the broken burs and review our experience with the removal of the broken burs in these cases.

Methods: All patients, who suffered the broken bur remained in the lower jaw due to the extraction of the impacted mandibular third molar and presented to our hospital from July 2019 to July 2024, were included in this retrospective study.

View Article and Find Full Text PDF

Objective: This study presents a novel digital interproximal enamel reduction (IER) clinical procedure, aiming to improve the effectiveness of IER processes in orthodontic treatment.

Methods: A malocclusion case of skeletal-class I and angle-class I was selected for the experimental investigation. A three-dimensional (3D) model of the dentition was constructed using scanning data from a plaster model.

View Article and Find Full Text PDF

Purpose: This systematic review evaluated the effect of different printing orientations on the physical-mechanical properties and accuracy of resin denture bases and related specimens.

Study Selection: Utilizing PRISMA 2020 guidelines, a comprehensive search of PubMed, Web of Science, Cochrane, and Scopus databases was conducted until June 2024. Included studies examined the accuracy, volumetric changes, and mechanical or physical properties of 3D-printed denture bases in various orientations.

View Article and Find Full Text PDF

The effect of the guide hole height on the accuracy of implant placement: an in-vitro study.

J Stomatol Oral Maxillofac Surg

January 2025

School of Dentistry, Chonnam National University, Gwangju, Republic of Korea; Department of Oral & Maxillofacial Surgery, Chonnam National University Hospital, Gwangju, Republic of Korea. Electronic address:

Objective: Guided implant surgery using three-dimensional (3-D) planning software and 3-D printed surgical guides has become a critical tool for enhancing accuracy. This study aims to determine the minimum guide hole height necessary to maintain implant placement accuracy.

Materials And Methods: Ten maxillary models with edentulous areas were created using CT and optical scan data.

View Article and Find Full Text PDF

The aim of this study was to assess the effect of a chlorhexidine digluconate solution (CHX) applied as an antiproteolytic agent for controlling erosive tooth wear or as part of the adhesive treatment on long-term bond strength to eroded dentin. Dentin specimens were abraded with a 600-grit silicon carbide (SiC) paper for 1 min (sound dentin - S), subsequently treated with 2% CHX for 1 min (with excess removed, followed by a 6-hour rest), and eroded by exposure to Coca-Cola for 5 min, three times a day, for 5 days (CHX-treated and eroded dentin - CHXE), or only eroded (eroded dentin - E). The specimens were acid-etched (15 s), rinsed (30 s), dried (15 s), and rehydrated with 1.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!