Resistance of internal-connection implant connectors under rotational fatigue loading.

Int J Oral Maxillofac Implants

Department of Prosthodontics School of Dentistry, University of Geneva, Switzerland.

Published: May 2007

AI Article Synopsis

  • This study examined the fatigue resistance of five different abutment connector configurations in dental implants, focusing on whether the antirotational mechanism affected this resistance.
  • The testing involved a rotating cantilever beam method to simulate real-life forces, aiming to identify load levels that led to failure in half of the specimens.
  • Results indicated that Easy abutments performed better in terms of fatigue resistance compared to ceramic and Multi-unit abutments, and the antirotational feature did not enhance mechanical strength.

Article Abstract

Purpose: To aid in developing mechanically optimized implant-abutment connectors, the fatigue resistance of 5 connector configurations of the Replace Select system (Easy abutment, Easy abutment without antirotational mechanism, Multi-unit abutment, Esthetic Alumina abutment, Esthetic Zirconia abutment) was investigated. Other purposes of the study were to determine whether the connector's antirotational mechanism participates in fatigue resistance and to compare the results with previous data on Straumann connectors.

Materials And Methods: The repetitive, alternating, and multivectorial intraoral force pattern was reproduced by subjecting the test specimens to the rotating cantilever beam test. To this end, the samples were spun around their long axis while clamped into a revolving collet on one end and loaded normal to their long axis on the other end. The aim was to determine the load level at which 50% of the specimens survived and 50% fractured before 106 cycles. Means were determined using the staircase procedure. They were fitted with 95% confidence intervals for intergroup comparisons.

Results: In the chosen testing configuration, 2 statistical groups emerged. The Easy abutments with and without antirotational mechanism were statistically similar, with mean failure loads in the 70 to 72 N range. Both ceramic and the Multi-unit abutments belonged to the second group, with mean failure loads in the 53 to 58 N range.

Conclusions: (1) The fatigue resistance of ceramic and the Multi-unit abutments was approximately 20% less than that of the Easy Abutments. (2) The antirotational mechanism did not participate in mechanical resistance. (3) The fatigue strength of the Easy abutment connectors was approximately 20% greater than the equivalent abutments in the Straumann system.

Download full-text PDF

Source

Publication Analysis

Top Keywords

antirotational mechanism
16
fatigue resistance
12
easy abutment
12
abutment esthetic
8
long axis
8
easy abutments
8
abutments antirotational
8
failure loads
8
ceramic multi-unit
8
multi-unit abutments
8

Similar Publications

Introduction: fixation of unstable intertrochanteric fractures presents a significant challenge, especially in the context of osteoporosis. Intramedullary implants have been established as superior to plate constructs. Our aim is to compare the complications and clinical outcomes of the Proximal Femur Nail (PFN) and Proximal Femur Nail Antirotation-2 (PFNA2) in managing unstable intertrochanteric fractures.

View Article and Find Full Text PDF

Intertrochanteric femur fracture is a common type of osteoporotic fracture in elderly patients, and postoperative femoral head varus following proximal femoral nail anti-rotation (PFNA) fixation is a crucial factor contributing to the deterioration of clinical outcomes. The cross-angle between the implant and bone might influence fixation stability. Although there is a wide range of adjustment in the direction of anti-rotation blades within the femoral neck, the impact of this direct variation on the risk of femoral head varus and its biomechanical mechanisms remain unexplored.

View Article and Find Full Text PDF

Changes in nail position and antirotation blade angles on the risk of femoral head varus in PFNA fixed patients: a clinical review and comprehensive biomechanical research.

Eur J Med Res

June 2024

Department of Orthopedics, Luzhou Key Laboratory of Orthopedic Disorders, The Affiliated Traditional Chinese Medicine Hospital, Southwest Medical University, No. 182, Chunhui Road, Luzhou, 646000, Sichuan Province, People's Republic of China.

Background: Femoral head varus triggers poor clinical prognosis in intertrochanteric fracture patients with proximal femoral nail antirotation (PFNA) fixation. Studies present that changes in nail position and screw insertion angles will affect fixation stability, but the biomechanical significance of these factors on the risk of femoral head varus has yet to be identified in PFNA fixed patients.

Methods: Clinical data in PFNA fixed intertrochanteric fracture patients have been reviewed, the relative position of intermedullary nail has been judged in the instant postoperative lateral radiography.

View Article and Find Full Text PDF

Background: Intertrochanteric fractures are highly prevalent among the elderly population, with approximately 90% occurring in individuals aged over 65. These geriatric fractures are associated with elevated mortality rates and significant functional impairment. Common treatment modalities for unstable intertrochanteric fractures include proximal femoral nail antirotation (PFNA) and the InterTan nail (IT).

View Article and Find Full Text PDF

The PFNA in treatment of intertrochanteric fractures with or without lateral wall fracture in elderly patients: a retrospective cohort study.

Eur J Med Res

September 2023

Trauma Medical Center, Department of Orthopaedic Surgery, West China Hospital, West China Medical School, Sichuan University, No. 37 Guoxue Road, Chengdu, 610041, Sichuan, People's Republic of China.

Background: There is no consensus about intertrochanteric fractures with lateral wall treated with intramedullary nail-proximal femoral nail antirotation (PFNA). The aim of the present study was to compare function outcomes between lateral wall and no lateral wall fractures after surgery by PFNA.

Methods: This retrospective study evaluated patients with or without lateral wall fractures who underwent PFNA between January 2015 and June 2018.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!