Objective: The aspiration of subglottic secretions colonized by bacteria pooled around the tracheal tube cuff due to inadvertent deflation (<20 cm H2O) of the cuff plays a relevant role in the pathogenesis of ventilator-associated pneumonia. We assessed the efficacy of an automatic, validated device for the continuous regulation of tracheal tube cuff pressure in preventing ventilator-associated pneumonia.

Design: Prospective randomized controlled trial.

Setting: Respiratory intensive care unit and general medical intensive care unit.

Patients: One hundred and forty-two mechanically ventilated patients (age, 64 +/- 17 yrs; Acute Physiology and Chronic Health Evaluation II score, 18 +/- 6) without pneumonia or aspiration at admission.

Interventions: Within 24 hrs of intubation, patients were randomly allocated to undergo continuous regulation of the cuff pressure with the automatic device (n = 73) or routine care of the cuff pressure (control group, n = 69). Patients remained in a semirecumbent position in bed.

Measurements And Main Results: The primary end point variable was the incidence of ventilator-associated pneumonia. Main causes for intubation were decreased consciousness (43, 30%) and exacerbation of chronic respiratory diseases (38, 27%). Cuff pressure <20 cm H2O was more frequently observed in the control than the automatic group (45.3 vs. 0.7% determinations, p < .001). However, the rate of ventilator-associated pneumonia with clinical criteria (16, 22% vs. 20, 29%) and microbiological confirmation (11, 15% vs. 10, 15%), the distribution of early and late onset, the causative microorganisms, and intensive care unit (20, 27% vs. 16, 23%) and hospital mortality (30, 41% vs. 23, 33%) were similar for the automatic and control groups, respectively.

Conclusions: Cuff pressure is better controlled with the automatic device. However, it did not result in additional benefits to the semirecumbent position in preventing ventilator-associated pneumonia.

Download full-text PDF

Source
http://dx.doi.org/10.1097/01.CCM.0000266686.95843.7DDOI Listing

Publication Analysis

Top Keywords

tracheal tube
8
tube cuff
8
automatic control
4
control tracheal
4
cuff pressure
4
pressure ventilated
4
ventilated patients
4
patients semirecumbent
4
semirecumbent position
4
position randomized
4

Similar Publications

Background: Emergency Front of Neck access eFONA) via cricothyroidotomy using a size 6 internal diameter tracheal tube is recommended by the Difficult Airway Society in the event of a 'can't intubate, can't oxygenate' (CICO) scenario in adults. There is a lack of clear guidance on whether to retain or remove a previously inserted supraglottic airway device (SAD) before eFONA. We aimed to study the effect of both neck extension and insertion of an SAD on sagittal cricothyroid membrane (CTM) height.

View Article and Find Full Text PDF

The incidence of arytenoid dislocation in abdominal surgery is relatively high, the cause is unknown, and it has not received sufficient attention. To identify the risk factors of arytenoid dislocation after abdominal surgery, and to establish a clinical prediction model based on relevant clinicopathological characteristics. We retrospectively collected the clinical data of 50 patients with arytenoid dislocation (AD) and 200 patients without AD after abdominal surgery with general anesthetic tracheal intubation in our Hospital from January 2013 to December 2019.

View Article and Find Full Text PDF

Ciprofol, a novel γ-aminobutyric acid receptor agonist, outperforms propofol with minimal cardiovascular effects, higher potency, reduced injection pain, and a broader safety margin. Despite these advantages, ciprofol's clinical research is still emerging. This study compares the median effective dose (ED) and adverse reactions of ciprofol and propofol, in conjunction with sufentanil, for suppressing cardiovascular responses during tracheal intubation.

View Article and Find Full Text PDF

Background: Airway management is a critical component of the care of patients experiencing cardiac arrest, but data from randomized trials on the use of video vs direct laryngoscopy for intubation in the setting of cardiac arrest are limited. Current AHA guidelines recommend placement of an endotracheal tube either during CPR or shortly after return of spontaneous circulation but do not provide guidance around intubation methods, including the choice of laryngoscope.

Research Question: Does use of video laryngoscopy improve the incidence of successful intubation on the first attempt, compared to use of direct laryngoscopy, among adults undergoing tracheal intubation after experiencing cardiac arrest?

Study Design And Methods: This secondary analysis of the Direct versus Video Laryngoscope (DEVICE) trial compared video laryngoscopy versus direct laryngoscopy in the subgroup of patients who were intubated following cardiac arrest.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!