Purpose: The objective of this study was to compare the diagnostic quality of low-dose computed tomography (CT) with that of standard-dose chest CT in the diagnosis of infectious lung diseases.
Materials And Methods: Thirty chest CT scans [high-resolution computed tomography (HRCT), 15; spiral CT, 15] were performed in HIV-positive patients with an infectious lung disease. Two additional slices were obtained at two lower exposures (HRCT, 120 kV/70 mAs and 120 kV/50 mAs; spiral CT, 100 kV/56 mAs and 100 kV/40 mAs) after chest routine CT. Observers compared the quality of the images obtained at different parameters and image noise. Objective evaluation of image noise was also made.
Results: Diagnostic image quality was excellent in 93% of the low-dose HRCT scans and in 86% of the low-dose spiral CT scans, rates that are always acceptable in any case. Significant differences were found in noise levels between the low-dose and reference scans; however, artifacts did not compromise detection of abnormalities. In HRCT, a mean reduction of 77% from the standard technique to the low-dose scan is possible in total and in lung effective doses. In spiral CT, this reduction is lower at 71%. These values can reach a further reduction with ultra-low-dose imaging (84% in HRCT and 80% in spiral CT).
Conclusions: Chest CT image quality appears to be adequate to evaluate pulmonary infectious diseases, even with an effective reduction in radiation dose. Standard-dose CT with a higher patient effective dose may be appropriate for selected cases.