Practical solution for control of the pre-analytical phase in decentralized clinical laboratories for meeting the requirements of the medical laboratory accreditation standard DIN EN ISO 15189.

Clin Lab

Uberörtliche Gemeinschaftspraxis für Laboratoriumsmedizin Leverkusen-Köln-Wenrath, Standort Köln, Labor Laser und Kollegen, Cologne, Germany.

Published: May 2007

This report was written in response to the article by Wood published recently in this journal. It describes a practical solution to the problems of controlling the pre-analytical phase in the clinical diagnostic laboratory. As an indicator of quality in the pre-analytical phase of sample processing, a target analyte was chosen which is sensitive to delay in centrifugation and/or analysis. The results of analyses of the samples sent by satellite medical practitioners were compared with those from an on-site hospital laboratory with a controllable optimized pre-analytical phase. The aim of the comparison was: (a) to identify those medical practices whose mean/median sample values significantly deviate from those of the control situation in the hospital laboratory due to the possible problems in the pre-analytical phase; (b) to aid these laboratories in the process of rectifying these problems. A Microsoft Excel-based Pre-Analytical Survey tool (PAS tool) has been developed which addresses the above mentioned problems. It has been tested on serum potassium which is known to be sensitive to delay and/or irregularities in sample treatment. The PAS tool has been shown to be one possibility for improving the quality of the analyses by identifying the sources of problems within the pre-analytical phase, thus allowing them to be rectified. Additionally, the PAS tool has an educational value and can also be adopted for use in other decentralized laboratories.

Download full-text PDF

Source

Publication Analysis

Top Keywords

pre-analytical phase
24
pas tool
12
practical solution
8
sensitive delay
8
hospital laboratory
8
problems pre-analytical
8
pre-analytical
7
phase
6
problems
5
solution control
4

Similar Publications

Thermal Desorption Hyphenated to Comprehensive Two-Dimensional Gas Chromatography-Time-of-Flight Mass Spectrometry for Trace Analysis in Raw Renewable Gases-Application to Hydrothermal Gasification.

J Sep Sci

January 2025

Department of Analytical, Bioanalytical Sciences and Miniaturization (LSABM) Chemistry, Biology and Innovation (CBI), UMR CNRS-ESPCI Paris, ESPCI Paris, PSL University, CNRS, Paris, France.

In the context of the energy transition, European countries pursue the common goal of increasing the share of renewable gases (from anaerobic digestion, pyrogasification, and hydrothermal gasification for instance) in the gas mix. Although produced gases are mainly composed of methane after upgrading, impurities of various natures and quantities may also be present in the produced raw gases and still after upgrading, including volatile organic compounds (VOCs) at trace levels that may have an impact on different stages of the gas chain even at low concentrations. These new renewable and/or low-carbon gases imply the need to develop new analytical tools to deeply characterize them, and thus fully manage their integration into the gas value chain.

View Article and Find Full Text PDF

Methotrexate (MTX) is an antifolic agent used in the first line of anti-inflammatory disease treatment and some oncologic issues. The metabolism of MTX is rapid, and the MTX concentration in the blood is not significant 24 h after administration. Unlike this, methotrexate polyglutamates (MTXPGs) can be informative biomarkers of drug exposure.

View Article and Find Full Text PDF

This study introduces an innovative bio-based sorbent bead crafted by integrating chitosan (CS) biopolymers, Fe(NO3)3 and polydopamine nanoparticles (PDA NPs) via glutaraldehyde crosslinking. The primary focus of this study was the concurrent separation of diverse tetracycline antibiotics (TCs), followed by rigorous reversed-phase liquid chromatography analysis. The fabricated CS/Fe@PDA sorbent beads were comprehensively characterized using scanning electron microscopy and energy-dispersive X-ray spectroscopy, revealing a surface rich in active carbon (C), nitrogen (N), and oxygen (O) moieties.

View Article and Find Full Text PDF

Background: Tavaborole (TAV), a benzoxaborole derivative, is an FDA-approved antifungal agent for treating onychomycosis, a common and persistent fungal infection of the toenails.

Objective: This study aimed to develop a robust stability-indicating HPTLC method to determine TAV in nanostructured lipid carriers (NLC) using a comprehensive approach that includes risk assessment, and Analytical Quality by Design.

Methods: The critical method parameters influencing the HPTLC results were screened using a Plackett-Burman screening design followed by its optimization using a central composite optimization design.

View Article and Find Full Text PDF

Solid Phase Microextraction-Gas Chromatography Triple Quadrupole Mass Spectrometry (SPME-GC-TQ/MS) was optimized and validated to specifically analyze aldehydes and furans after drying by conventional as well as modern pre-drying technique i.e. pulsed electric field (PEF).

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!