Background: Monitoring of the electrocardiogram (ECG) in premature infants with conventional adhesive-backed electrodes can harm their sensitive skin. Use of an electrode belt prevents skin irritation, but the effect of belt pressure on respiratory function is unknown. A strain gauge sensor is described which measures applied belt tension.

Method: The device frame was comprised of an aluminum housing and slide to minimize the device weight. Velcro tabs connected housing and slide to opposite tabs located at the electrode belt ends. The slide was connected to a leaf spring, to which were bonded two piezoresistive transducers in a half-bridge circuit configuration. The device was tested for linearity and calibrated. The effect on infant respiratory function of constant belt tension in the normal range (30 g-90 g) was determined.

Results: The mechanical response to a step input was second order (fn = 401 Hz, zeta = 0.08). The relationship between applied tension and output voltage was linear in the range 25-225 gm of applied tension (r2 = 0.99). Measured device sensitivity was 2.18 mV/gm tension using a 5 V bridge excitation voltage. When belt tension was increased in the normal range from 30 gm to 90 gm, there was no significant change in heart rate and most respiratory functions during monitoring. At an intermediate level of tension of 50 gm, pulmonary resistance and work of breathing significantly decreased.

Conclusion: The mechanical and electrical design of a device for monitoring electrocardiogram electrode belt tension is described. Within the typical range of application tension, cardiovascular and respiratory function are not substantially negatively affected by electrode belt force.

Download full-text PDF

Source
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC1868740PMC
http://dx.doi.org/10.1186/1475-925X-6-13DOI Listing

Publication Analysis

Top Keywords

belt tension
16
respiratory function
16
electrode belt
16
monitoring electrocardiogram
12
belt
9
tension
9
premature infants
8
housing slide
8
normal range
8
applied tension
8

Similar Publications

A Case Study on the Possibility of Extending the Service Life of the Demining Machine Belt.

Materials (Basel)

October 2024

Department of Technological Engineering, Faculty of Mechanical Engineering, University of Žilina, Univerzitná 8215/1, 010 26 Žilina, Slovakia.

The operational practice of the design of the Bozena 5 demining machine has shown that its belts are the critical component that fundamentally affects the functionality of the entire machine. This article is a practical continuation and extension of the previous research results from the point of view of materials (research of the uniaxial fatigue life in bending and torsion), calculation (creation of the necessary mathematical, analytical and numerical models for the research) and construction (i.e.

View Article and Find Full Text PDF

Objective: The goal of this study was to examine the relationship between lap belt tension and force measured at the iliac wing and the effects of model type and torso posture on this relationship. From this analysis, preliminary transfer functions were developed to predict loads applied to the iliac wing as a function of lap belt tension at magnitudes typically measured in sled and vehicle crash tests.

Methods: A DOE study was conducted to provide a robust assessment of the lap belt-pelvis load relationship under various conditions.

View Article and Find Full Text PDF

Objectives: Elevated head injury incidence in infants compared to toddlers involved as occupants in motor vehicle crashes has been demonstrated in multiple population-representative crash databases. Further, experimental studies have revealed a potential injury mechanism impact between a rear-facing, CRS-restrained child and the back of the vehicle seat or console on the row in front of the CRS. Subsequently, experimental studies have suggested that bracing the CRS against the seat immediately in front of the CRS could mitigate head injury, but also indicated that more research was necessary.

View Article and Find Full Text PDF

Reliability and validity of belt-stabilized and tension dynamometry for assessing hip strength and power in uninjured adults.

Phys Ther Sport

September 2024

School of Exercise and Rehabilitation Sciences, Health & Human Services, University of Toledo, Mail Stop 1199, 201 W. Bancroft St, Toledo, OH, 43606-3390, USA. Electronic address:

Objectives: To investigate the intra-rater reliability and validity of belt-stabilized and tension dynamometry to assess hip muscle strength and power.

Design: Repeated measures.

Setting: Biomechanics laboratory.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!