Purpose: Reduction of radiation exposure at an adequate image quality by optimizing the radiation quality for a new system of full-field digital mammography using a digital detector (a-Se).
Materials And Methods: The investigations were performed using a digital mammography system Novation (Siemens, Erlangen). The system was constructed with a bimetal anode (molybdenum and tungsten) and the possibility of changing the filter (molybdenum/rhodium). The test object was the Wisconsin Mammography Random Phantom Model 152 A (Radiation Measurements Inc.) of which images were acquired using the digital technique with the tungsten anode and rhodium filter at different tube voltages (26-35 kV) and tube loads (40-100 mAs) and compared to images in the molybdan/molybdan molybdenum/molybdenum technique. To quantify the image quality, we used the detection rate of the simulated lesions in the phantom.
Results: Increasing the tube voltage significantly decreases the average glandular dose when using AEC (Automatic Exposure Control), i. e., constant detector dose. At the same time, the image quality decreases significantly with respect to the detection rate (26 kV, 1 mGy, 95.1 %; 35 kV, 0.7 mGy, 82.7 %). As a good compromise between the necessary diagnostic image quality and the lowest dose exposition, 28 kV and 60 mAs were selected for imaging with the tungsten/rhodium anode/filter combination. A further change to the tube load did not make sense because a decrease of 10 % resulted in a significant decrease in the detection rate while only a 2 % increase in detection rate was achieved for a 65 % increase in radiation exposure.
Conclusion: The results of this phantom study demonstrate that the routine use of the tungsten anode in combination with a rhodium filter for full-field digital mammography with an a-Se detector in contrast to a molybdan/molybdan molybdenum/molybdenum anode/filter combination results in a reduction of the average glandular dose of up to 30 % without loss of diagnostic image quality.
Download full-text PDF |
Source |
---|---|
http://dx.doi.org/10.1055/s-2007-962983 | DOI Listing |
J Orthop Surg Res
January 2025
Excellence Center for Hip & Knee Arthroplasty, Department of Orthopedic Surgery, Zuyderland Medical Center, Heerlen, The Netherlands.
Introduction: In 2020, 368 million people globally were affected by knee osteoarthritis, and prevalence is projected to increase with 74% by 2050. Relatively high rates of dissatisfactory results after total knee arthroplasty (TKA), as reported by approximately 20% of patients, may be caused by sub-optimal knee alignment and balancing. While mechanical alignment has traditionally been the goal, patient-specific alignment strategies are gaining interest.
View Article and Find Full Text PDFBackground: The quality of life for low vision patients is a comprehensive concept that involves multiple aspects, including physical, psychological, and social aspects. Visual impairment is an important factor in the decline of patients' quality of life. The purpose of this study is to investigate the impact and feasibility of carrying out low vision rehabilitation nursing on the quality of life and nursing burden of low vision patients.
View Article and Find Full Text PDFEur J Hum Genet
January 2025
Institute for Genomic Statistics and Bioinformatics, Bonn, NRW, Germany.
The facial gestalt (overall facial morphology) is a characteristic clinical feature in many genetic disorders that is often essential for suspecting and establishing a specific diagnosis. Therefore, publishing images of individuals affected by pathogenic variants in disease-associated genes has been an important part of scientific communication. Furthermore, medical imaging data is also crucial for teaching and training deep-learning models such as GestaltMatcher.
View Article and Find Full Text PDFAesthetic Plast Surg
January 2025
Plastic, Reconstructive and Aesthetic Surgery Department, Fondazione Policlinico Universitario Campus Bio-Medico, Via Alvaro del Portillo, 200, 00128, Roma, Italy.
Introduction: Renuvion was the first FDA approved helium plasma device utilized for subdermal tissue heating to reduce skin laxity. The purpose of this study was to demonstrate that the use of Renuvion improves the outcomes, skin quality and reduces the edema faster after lipoabdominoplasty.
Materials And Methods: Patients with abdominal skin laxity after a weight loss of at least 20 kg, nonsmokers, without major comorbidities, with a minimum 2-year follow-up and standardized pre- and postoperative photographs were included in this study.
Sci Rep
January 2025
Hive AI Innovation Studio, Department of Computer Science and Engineering, University of Louisville, Louisville, KY, 40292, USA.
Nailfold Capillaroscopy (NFC) is a simple, non-invasive diagnostic tool used to detect microvascular changes in nailfold. Chronic pathological changes associated with a wide range of systemic diseases, such as diabetes, cardiovascular disorders, and rheumatological conditions like systemic sclerosis, can manifest as observable microvascular changes in the terminal capillaries of nailfolds. The current gold standard relies on experts performing manual evaluations, which is an exhaustive time-intensive, and subjective process.
View Article and Find Full Text PDFEnter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!