Purpose: Craniospinal irradiation (CSI) has some geometric uncertainties, especially at the junction zone. In this study we tried to evaluate how possible random setup errors of the collimator angle may contribute to these uncertainties.

Materials And Methods: Cranial and spinal fields were drawn on RW3 solid water phantom in accordance with the divergence matching technique (DMT). Field dimensions were 18x18 cm and 6x30 cm, respectively. We placed light-insulated Kodak X-Omat V films at the junction zone, then we irradiated the films with different collimator angles with both Co-60 and 4 MV conditions, and determined how the junction zone was affected from random setup errors in DMT.

Results: 10.6 degrees collimator angle was proper for 30 cm upper spinal field. For Co-60 machine the dose homogeneity of this angle was +4.5%. For the angles of 8, 9, 11 and 12 degrees the homogeneities were -13%, - 11%, +5% and +10 %, respectively. For 4 MV photon the dose homogeneity of the 10.6 degrees collimator angle was +3%. For the angles of 8, 9, 11 and 12 degrees the homogeneities were -17%, -14%, +6 % and +13%, respectively.

Conclusion: As the CSI has some geometric uncertainties, serious dose inhomogeneities may occur at the junction zone. The collimator angle is of great importance and any random setup errors may not be tolerated.

Download full-text PDF

Source

Publication Analysis

Top Keywords

collimator angle
20
junction zone
20
setup errors
16
random setup
12
csi geometric
8
geometric uncertainties
8
106 degrees
8
degrees collimator
8
dose homogeneity
8
angles degrees
8

Similar Publications

All-angle unidirectional flat-band acoustic metasurfaces.

Nat Commun

January 2025

Department of Electrical and Computer Engineering, National University of Singapore, Singapore, Singapore.

Flat bands have empowered novel phenomena such as robust canalization with strong localization, high-collimation and low-loss propagation. However, the spatial symmetry protection in photonic or acoustic lattices naturally forces flat bands to manifest in pairs aligned at an inherently specific angle, resulting in a fixed bidirectional canalization. Here, we report an acoustic flat-band metasurface, allowing not only unidirectional canalization at all in-plane angles but also robust tunability in band alignment.

View Article and Find Full Text PDF

Excessive kidney mobility is an underestimating challenge for surgeons during retrograde intrarenal surgery (RIRS) and extracorporeal shock wave lithotripsy (ESL). There is no technique approved as a gold standard procedure for reducing excessive kidney mobility. The study aimed to uncover predictive factors for determining excessive renal mobility by utilizing clinicodemographic characteristics and noncontrast computed tomography (NCCT) data.

View Article and Find Full Text PDF

Fiber-Optic Photoacoustic Gas Microprobe Based on Linear Spot-Type Multipass Cell.

Anal Chem

January 2025

School of Optoelectronic Engineering and Instrumentation Science, Dalian University of Technology, Dalian, Liaoning 116024, China.

A linear spot-type multipass cell-enhanced fiber-optic photoacoustic gas microprobe is proposed. To further reduce the volume of the gas chamber and enhance the photoacoustic signal, we designed the cross section of the photoacoustic tube as a slit with a height of 10 mm and a width of 1.5 mm.

View Article and Find Full Text PDF

Background: High-dose-rate (HDR) brachytherapy using Iridium-192 as a radiation source is widely employed in cancer treatment to deliver concentrated radiation doses while minimizing normal tissue exposure. In this treatment, the precision with which the sealed radioisotope source is delivered significantly impacts clinical outcomes.

Purpose: This study aims to evaluate the feasibility of a new four-dimensional (4D) in vivo source tracking and treatment verification system for HDR brachytherapy using a patient-specific approach.

View Article and Find Full Text PDF

Quality assurance practices performed before treatment are believed to identify various potential errors. In this study, 2-dimensional (2D) dosimetric results were analyzed by making some intentional mistakes in six different treatment plans. In this way, the detectability of errors was investigated.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!