Unlabelled: Several studies have described nonuniform blurring of myocardial perfusion imaging (MPI) due to respiration. This article describes a technique for correcting the respiration effect and assesses its effectiveness in clinical studies.

Methods: Simulated phantoms, physical phantoms, and patient scans were used in this study. A heart phantom, which oscillated back and forth, was used to simulate respiration. The motion was measured on a gamma-camera supporting list-mode functionality synchronized with an external respiratory strap or resistor sensor. Eight clinical scans were performed using a 1-d (99m)Tc-sestamibi protocol while recording the respiratory signal. The list-mode capability along with the strap or sensor signals was used to generate respiratory bin projection sets. A segmentation process was used to detect the shift between the respiratory bins. This shift was further projected to the acquired projection images for correction of the respiratory motion. The process was applied to the phantom and patient studies, and the rate of success of the correction was assessed using the conventional bull's eye maps.

Results: The algorithm provided a good correction for the phantom studies. The shift after the correction, measured by a fitted ellipsoid, was <1 mm in the axial direction. The average motion due to respiration in the clinical studies was 9.1 mm in the axial direction. The average shift between the respiratory phases was reduced to 0.5 mm after correction. The maximal change in the bull's eye map for the clinical scans after the correction was 6%, with a mean of 3.75%. The postcorrection clinical summed perfusion images were more uniform, consistent, and, for some patients, clinically significant when compared with the images before correction for respiration.

Conclusion: Myocardial motion generated by respiration during MPI SPECT affects perfusion image quality and accuracy. Motion introduced by respiration can be corrected using the proposed method. The degree of correction depends on the patient respiratory pattern and can be of clinical significance in certain cases.

Download full-text PDF

Source
http://dx.doi.org/10.2967/jnumed.106.037390DOI Listing

Publication Analysis

Top Keywords

myocardial perfusion
8
respiratory
6
correction
5
correction heart
4
heart motion
4
respiration
4
motion respiration
4
respiration clinical
4
clinical myocardial
4
perfusion spect
4

Similar Publications

Cardiac Positron Emission Tomography (PET) can be used for the assessment of myocardial perfusion. Compared to other cardiac imaging techniques, notably Single Photon Emission Computer Tomography (SPECT), cardiac PET offers superior image resolution, higher accuracy, quantitative measures of myocardial perfusion, lower radiation exposure, and shorter image acquisition time. However, PET tends to be costlier and less widely available than SPECT due to the specialized equipment needed for generating the necessary radiotracers.

View Article and Find Full Text PDF

Introduction: ST-elevation myocardial infarction (STEMI) is one of the most prevalent presentations in young patients. It is essential to emphasise that each minute of delay in providing medical care is negatively correlated to the patient's prognosis. The present study was carried out to evaluate the ischaemia-reperfusion times in patients ≤40 years of age versus individuals >40 years of age and their association with mortality and major adverse cardiac event (MACE) over the long term.

View Article and Find Full Text PDF

Ventricular arrhythmias induced by ischemia/reperfusion injury limits the therapeutic effect of early reperfusion therapy for acute myocardial infarction. This study investigated the protective effects of the β2-adrenergic receptor (β2-AR) agonist clenbuterol against ischemia/reperfusion-induced arrhythmias and the underlying mechanism. Anesthetized rats were subjected to 10-min left coronary artery occlusion and 10-min reperfusion in vivo.

View Article and Find Full Text PDF

Background: The fluorescent dye indocyanine green (ICG) has been used to identify anatomical structures intraoperatively in coronary artery bypass grafting (CABG). This study aimed to evaluate the feasibility of using ICG to assess graft patency and territorial distribution of myocardial reperfusion during CABG.

Methods: Porcine arrested hearts (n = 18) were used to evaluate territorial distribution of native coronary arteries and of a coronary bypass constructed with porcine saphenous vein graft (SVG) using ICG.

View Article and Find Full Text PDF