This study reviewed 863 alerts generated from the infusion of anticoagulants in 355 patients from October 2003 to January 2005. Alerts were generated by smart infusion technology pumps and recorded in the devices' memory. The most common alerts were underdose alerts (59.8%), followed by overdose alerts (31.3%) and duplicate drug therapy alerts (8.9%). In response to the alerts, users' most frequent action was to cancel (46.5%) or reprogram (43.1%) the infusions. The highest percentage of alerts occurred from 2 to 4 p.m. During the study, there were 4 infusion rate errors, compared with 15 in the immediately preceding 16-month period. In conclusion, smart infusion technology intercepted keypad entry errors, thereby reducing the likelihood of intravenous anticoagulant overdose or underdose. Dose or infusion rate programming during intravenous anticoagulation is an important targets for medication safety interventions.
Download full-text PDF |
Source |
---|---|
http://dx.doi.org/10.1016/j.amjcard.2006.10.069 | DOI Listing |
Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!