Cyanide is a regulated contaminant in drinking water in the United States. This paper describes an ion chromatography method with pulsed amperometric detection (PAD) that directly determines free cyanide in drinking water. Samples are treated with sodium hydroxide to stabilize cyanide and with a cation-exchange cartridge to remove transition metals. Cyanide is separated by anion-exchange chromatography and detected by PAD with a waveform optimized for cyanide and used with a disposable silver working electrode. The recovery of cyanide spiked into five water samples was >80%. With an MDL of 1.0 microg/L, this method determines cyanide concentrations well below the reporting limits for free cyanide in drinking water.

Download full-text PDF

Source
http://dx.doi.org/10.1016/j.chroma.2007.02.083DOI Listing

Publication Analysis

Top Keywords

drinking water
16
free cyanide
12
cyanide drinking
12
cyanide
9
ion chromatography
8
pulsed amperometric
8
amperometric detection
8
water samples
8
water
5
direct determination
4

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!