Objectives: To observe the effect of both acute and medium-term magnetic stimulation of the pelvic floor on detrusor function amongst women with idiopathic detrusor overactivity (IDO).

Methods: Two separate studies were undertaken amongst women with a sole diagnosis of IDO. The first study assessed the acute effect of magnetic stimulation (provided by Neocontrol ) on detrusor function during the filling phase of standard cystometry. Multiple filling cycles were performed with stimulation at a different key moment in each. This was done to establish that the device could influence the detrusor. Subsequently, a randomized sham control trial was performed to assess clinical efficacy. A total of 20 treatments, each of 20 minutes duration, were administered over six weeks with follow-up six weeks thereafter. Half the patients received therapy from a genuine device, the others receiving fake treatment on an identical looking/sounding sham device. The sham device contained a deflector plate to degrade the magnetic field and was located in a separate room. Outcome measures included changes in a 24 hour fluid volume chart, urine loss (24 hour pad test) and quality of life instruments.

Results: Amongst 10 patients receiving stimulation during cystometry, volume at first involuntary detrusor contraction during filling rose from a median value of 240ml (Inter-quartile range (IQR) 210-300) to 285ml (IQR 231-320), p = 0.03 and maximum detrusor pressure decreased from 40cm water (IQR 34-45) to 33cm water (IQR 25-41), p<0.01. The RCT was completed by 29 of 44 (66%) recruits. Of these, 15 of 29 (52%) received active treatment and 14 of 29 (48%) sham therapy. Active therapy significantly reduced the number of urge episodes per day, p<0.01. With respect to baseline, actively treated patients experienced significant reduction in voids per day and quality of life but this trend did not reach significance when compared to the sham group, partly due to unexpected difficulty in recruitment which yielded an underpowered sample size for these outcome measures.

Conclusions: Magnetic stimulation reduces detrusor contractility in the acute phase of administration. Although the treatment was well tolerated and urge episodes reduced following prolonged therapy, no statistically significant improvement was observed in quality of life indices or measured 24 hour urinary loss. The treatment cannot be recommended for women with IDO.

Download full-text PDF

Source
http://dx.doi.org/10.1016/j.eururo.2007.02.026DOI Listing

Publication Analysis

Top Keywords

magnetic stimulation
12
women idiopathic
8
idiopathic detrusor
8
detrusor overactivity
8
randomized sham
8
detrusor function
8
sham device
8
water iqr
8
detrusor
7
stimulation
5

Similar Publications

Enzyme-linked immunosorbent spot analysis is frequently used to investigate immune responsiveness during clinical trials. However, ELISpot classically utilizes peripheral blood mononuclear cell isolates from whole blood, requiring relatively high blood draw volumes and removing both granulocytes and bound drug. Here, we describe a novel protocol whereby CD45 cells are magnetically isolated from human whole blood and co-incubated with serum isolated from the same subject.

View Article and Find Full Text PDF

SARS-CoV-2 infection could trigger autoimmune disease. We report a case of concomitant exacerbation of Graves orbitopathy (GO) and myasthenia gravis (MG) after SARS-CoV-2 infection. A 43-year-old woman had diplopia, proptosis, and swollen eyelids.

View Article and Find Full Text PDF

Introduction: Corticotropin releasing hormone (CRH)-stimulated bilateral inferior petrosal sinus sampling (BIPSS) is the most accurate procedure in the differential diagnosis of adrenocorticotropic hormone (ACTH)-dependent Cushing's syndrome (CS) with a sensitivity of 88-100% and a specificity of 67-100%. However, CRH is not available globally currently. We undertook this study of BIPSS using lysine vasopressin (LVP) as an agent to stimulate the release of ACTH from corticotrophs.

View Article and Find Full Text PDF

Transcranial ultrasound stimulation (TUS) presents challenges in ultrasound wave transmission through the skull, affecting study outcomes due to aberration and attenuation. While planning strategies incorporating 3D computed tomography (CT) scans help mitigate these issues, they expose participants to radiation, which can raise ethical concerns. A solution involves generating skull masks from participants' anatomical magnetic resonance imaging (MRI).

View Article and Find Full Text PDF

Background: Late-life depression (LLD) is often accompanied by cognitive impairment, which may persist despite antidepressant treatment. Repetitive transcranial magnetic stimulation (rTMS) is an efficacious treatment for depression, with potential benefits on cognitive functioning. However, research on cognitive effects is inconclusive, relatively sparse in LLD, and predominantly focused on group-level cognitive changes.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!