Results of polyaxial locked-plate fixation of periarticular fractures of the knee.

J Bone Joint Surg Am

Orthopaedic Trauma Service, Florida Orthopedic Institute, 13020 Telecom Parkway, Temple Terrace, FL 33637, USA.

Published: March 2007

Background: Locked-plate fracture-fixation techniques and designs continue to evolve. Polyaxial locking plates that allow screw angulation and end-point locking have become available; however, there are no clinical data documenting their strength and efficacy, to our knowledge. The purpose of this study was to evaluate the clinical performance of a variable-axis locking plate in a multicenter series of periarticular fractures about the knee.

Methods: Between 2003 and 2005, fifty-four patients with a total of fifty-six fractures were treated with a polyaxial locked-plate fixation system (DePuy, Warsaw, Indiana). There were twenty male patients and thirty-four female patients with a mean age of fifty-seven years. There were twenty-five distal femoral fractures and thirty-one proximal tibial fractures. Twelve of the fractures were open. Clinical and radiographic data, including changes in alignment, hardware breakage, or other mechanical complications of the device, were retrospectively reviewed. Function was assessed with use of the Knee Society scores. One patient with a bilateral fracture died less than three months postoperatively, and two patients were lost to follow-up prior to union. Fifty-two fractures in fifty-one patients were followed to union or for a minimum of six months; the mean duration of follow-up was nine months (range, six to twenty-five months).

Results: Forty-nine (94%) of the fifty-two fractures united. There were no mechanical complications. Most importantly, there was no evidence of varus collapse as a result of polyaxial screw failure. There were three deep infections and one aseptic nonunion. No plate fractured, and no screw cut out.

Conclusions: The variable-axis locking plates performed well, with a high rate of fracture union and no evidence of varus collapse due to failure of the polyaxial screw fixation, in a series of complex fractures about the knee. Complication rates were similar to those for historical controls treated with fixed-trajectory locking plates. Polyaxial locking plates offer more fixation versatility without an apparent increase in mechanical complications or loss of reduction.

Download full-text PDF

Source
http://dx.doi.org/10.2106/JBJS.F.00510DOI Listing

Publication Analysis

Top Keywords

locking plates
16
mechanical complications
12
fractures
9
polyaxial locked-plate
8
locked-plate fixation
8
periarticular fractures
8
fractures knee
8
polyaxial locking
8
variable-axis locking
8
fifty-two fractures
8

Similar Publications

Treatment of anterior rim fractures of the distal radius with an anterior rim plate: A retrospective study.

J Hand Surg Eur Vol

January 2025

Department of Orthopedics, E-Da Hospital, I-Shou University/School of Medicine, College of Medicine, I-Shou University, Kaohsiung, Taiwan.

Treatment of distal radial fractures that include small anterior rim fragments can be difficult. We retrospectively reviewed 19 patients in whom an anterior rim plate with locking screws was used. After a median follow-up of 18 months (range 6-32; interquartile range (IQR) 14, 26), the median wrist flexion and extension arc was 70° (range 50-80; IQR 60, 70), the median grip strength was 80% of the contralateral side (range 52-104; IQR 77, 88), the median visual analogue scale score for pain was 0 (range 0-5; IQR 0, 1), the median disabilities of the arm, shoulder and hand score was 2 (range 0-59; IQR 0, 11) and the median modified Mayo wrist score was 80 (range 35-100; IQR 75, 85).

View Article and Find Full Text PDF

[Outcomes of Retrograde Femoral Nail Osteosynthesis of Intraarticular Fractures of the Distal Femur].

Acta Chir Orthop Traumatol Cech

January 2025

Klinika ortopedie a traumatologie pohybového ústrojí Fakultní nemocnice Plzeň.

Purpose Of The Study: Intraarticular fractures of the distal femur rank among the most severe musculoskeletal injuries. Various treatment options, such as plate osteosynthesis or retrograde nailing, can be employed. This study aims to evaluate the clinical outcomes and complications of intraarticular distal femoral fractures treated with retrograde femoral nail, with particular emphasis on C3 fractures.

View Article and Find Full Text PDF

Distal augmentation in unstable distal clavicle fractures: a retrospective cohort study of 101 cases.

Arch Orthop Trauma Surg

January 2025

The Department of Orthopaedic Surgery, Kaohsiung Chang Gung Memorial Hospital, Chang Gung University College of Medicine, 123, Ta Pei Road, Niao Sung Dist, Kaohsiung, Taiwan.

Introduction: The optimal management strategy for unstable distal clavicular fractures remains controversial. Recent studies on plate techniques have reported good-to-excellent outcomes with no serious complications. The questions are that: (1) Does the use of wire augmentation with locking plate in distal part (distal wire augmentation) reduce radiographic loss of reduction (RLOR) and get earlier bony union in distal clavicular fractures? (2) Which fixation methods are associated with a higher incidence of acromioclavicular (AC) joints arthritis or subluxation? We collected and analyzed clinical studies on different plate fixation methods for unstable fractures to identify the best surgical modality.

View Article and Find Full Text PDF

Background: Isolated distal fibula fractures (DFF) are usually treated with open reduction and internal fixation (ORIF) and non-weight-bearing protocols. The study assessed the outcomes of immediate weight-bearing on DFF healing and stability after lateral locking plating.

Materials And Methods: For this study, 49 patients affected by isolated DFF were enrolled.

View Article and Find Full Text PDF

Medial opening wedge high tibial osteotomy (HTO) treats medial knee osteoarthritis by realigning the knee joint, though it still carries quite a high risk of complications. A new Variable Fixation Locking Screw technology, designed to gradually reduce construct stiffness and promote bone healing, aims to address these issues. This observational study evaluates the safety and effectiveness of this innovative approach in improving clinical outcomes.

View Article and Find Full Text PDF