Evaluation of the mechanical strength and patency of functional end-to-end anastomoses.

Surg Endosc

Division of Gastroenterological Surgery, Department of Clinical Molecular Medicine, Kobe University Graduate School of Medicine, 7-5-2, Kusunoki-cho, Chuo-ku, Kobe, 650-0017, Japan.

Published: September 2007

Background: Although many mechanical devices are being developed for functional end-to-end anastomosis (FETEA), the best choice remains to be determined. The mechanical strength of each device is an important factor associated with anastomotic leakage. In turn, anastomotic patency is related to postoperative stenosis. This study examined the bursting pressure and circumference of several FETEAs.

Methods: Young domestic pigs were used in this study. In experiment 1, a 10-cm segment of intestine was transected using an EndoGIA 60 blue, EndoGIA 60 green, or GIA 60 blue stapler, and the bursting pressure of each instrument's mechanical staple line was examined. In experiment 2, after FETEA had been created with EndoGIA 60 blue, the bursting sites were examined, and the bursting pressures of buttressed and nonbuttressed FETEA were measured and compared. In experiment 3, the circumference of FETEA was compared between the closed technique and reinforcement.

Results: The bursting pressure of EndoGIA 60 blue was significantly higher than that of the other devices. When an anastomotic crotch was buttressed, the bursting pressure (44 +/- 13 mmHg) was significantly increased over that of the nonbuttressed FETEA (27.5 +/- 5.8 mmHg; p < 0.05). The circumference of FETEA using the wide technique (92.7 +/- 2.3 mm) was larger than that using the offset technique (55.7 +/- 5.8 mm).

Conclusions: The bursting pressure of the mechanical staple line using the EndoGIA 60 stapler was the strongest. Functional end-to-end anastomosis was stronger when created with buttress of an anastomotic crotch, and larger circumferences were created with the wide technique.

Download full-text PDF

Source
http://dx.doi.org/10.1007/s00464-006-9131-6DOI Listing

Publication Analysis

Top Keywords

bursting pressure
20
functional end-to-end
12
endogia blue
12
mechanical strength
8
end-to-end anastomosis
8
examined bursting
8
mechanical staple
8
nonbuttressed fetea
8
circumference fetea
8
anastomotic crotch
8

Similar Publications

Background: Intestinal ischemia/reperfusion (I/R) injury can occur in a wide variety of diseases and surgeries. If necessary, the blood flow should be restored, including re-anastomosis by removing the intestines with impaired circulation. In this process, anastomotic strength is as important as inflammatory responses and oxidative stress.

View Article and Find Full Text PDF

Effective management of nonconvulsive status epilepticus following cardiac surgery: a case report.

Gen Thorac Cardiovasc Surg Cases

January 2025

Department of Cardiovascular Surgery, Japan Organization of Occupational Health and Safety, Osaka Rosai Hospital, Sakai, Osaka, 591-8025, Japan.

Background: Epileptic seizures following adult cardiovascular surgery occur in 0.9-3% of patients, with the condition in 3-12% of these patients progressing to status epilepticus (SE). SE is a severe condition that significantly impacts prognosis and necessitates early diagnosis and treatment.

View Article and Find Full Text PDF

Cardiovascular diseases (CVDs) were responsible for approximately 19 million deaths in 2020, marking an increase of 18.7% since 2010. Biological decellularized patches are common therapeutic solutions for CVD such as cardiac and valve defects.

View Article and Find Full Text PDF

The purpose of this study was to clarify sex differences in the inhibition of sympathetic vasomotor outflow which is caused by the loading of cardiopulmonary baroreceptors. Ten young males and ten age-matched females participated. The participants underwent a passive leg raising (PLR) test wherein they were positioned supine (baseline, 0º), and their lower limbs were lifted passively at 10º, 20º, 30º, and 40º.

View Article and Find Full Text PDF

Objective: To evaluate the sparing effects of fentanyl and maropitant on sevoflurane minimum alveolar concentrations that block autonomic responses (MACBAR) and the hemodynamic and electroencephalographic responses to noxious stimuli in dogs.

Methods: The sevoflurane MACBAR was determined in 5 healthy male Beagles with or without continuous infusions of fentanyl and maropitant. Then, intermittent noxious stimulation was applied at 1.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!