Progress towards a high-resolution retinal prosthesis.

Conf Proc IEEE Eng Med Biol Soc

Doheny Retina Institute, Doheny Eye Institute, Department of Ophthalmology, Keck School of Medicine, University of Southern California, Los Angeles, CA.

Published: October 2012

Simulations of artificial vision suggest that 1000 electrodes may be required to restore vision to individuals with diseases of the outer retina. In order to achieve such an implant, new technology is needed, since the state-of-the-art implantable neural stimulator has at most 22 contacts with neural tissue. Considerable progress has been made towards that goal with the development of image processing, microelectronics, and polymer based electrodes and interconnects. An image processing system has been realized that is capable of real-time implementation of image decimation and filtering (for example, edge detection). Application specific integrated circuits (ASICs) have been designed and tested to demonstrate closed loop power control and efficient microstimulation. A novel packaging process has been developed that is capable of simultaneously forming a receiver coil, interconnects, and stimulating electrodes.

Download full-text PDF

Source
http://dx.doi.org/10.1109/IEMBS.2005.1616215DOI Listing

Publication Analysis

Top Keywords

image processing
8
progress high-resolution
4
high-resolution retinal
4
retinal prosthesis
4
prosthesis simulations
4
simulations artificial
4
artificial vision
4
vision 1000
4
1000 electrodes
4
electrodes required
4

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!