Arthrosopic knot pushers. Does one size fit all?

Int J Clin Pract

Oswestry Rotational, Baschurch, Shropshire, UK.

Published: May 2008

Little is known about how knot-pusher design affects arthroscopic knot tying. In our practice, we observed the knot-pusher riding onto the arthroscopic knot at the point of maximum tightening. This can lead to snagging of the knot, which is undesirable as it may lead to loosening of, or damage to the knot. The aim of this study is to determine the optimum size of a knot-pusher to efficiently push the knot without overriding or snagging it. We used an apparatus to model arthropcopic knot tying. Ten examples each of the Duncan loop were tied under controlled conditions of load using one polydioxanone (PDS) monofilament absorbable suture (Ethicon, Livingston, UK), two Ethibond, two Fibrewire and two Panacryl. The loop of the knot was then secured and a 50 N force applied to tension the knot. The suture diameter was measured. Then the knot diameter was measured in two planes using an analogue micrometer. The internal diameter of a Mitek knot-pusher was measured. The mean maximum diameter for each knot was respectively PDS, 2.061 +/- 0.13 mm; Panacryl, 1.907 +/- 0.14 mm; Ethibond, 1.717 +/- 0.16 mm and Fibrewire, 1.654 +/- 0.14 mm. There were significant differences in size between knots tied with different materials except between Ethibond and Fibrewire where the difference was not significant. For each set of knots the smallest maximum knot diameter observed was identified. This was respectively PDS, 1.92 mm; Ethibond, 1.476 mm; Fibrewire, 1.488 mm and Panacryl, 1.715 mm. The internal diameter of a Mitek knot-pusher was found to be 1.95 mm. The current Mitek knot-pusher appears to be well suited to one PDS and two Panacryl. It appears less ideal for two Ethibond and two Fibrewire. One knot-pusher does not fit all and we suggest that different knot-pushers be used for different suture materials.

Download full-text PDF

Source
http://dx.doi.org/10.1111/j.1742-1241.2006.01263.xDOI Listing

Publication Analysis

Top Keywords

ethibond fibrewire
12
mitek knot-pusher
12
knot
11
arthroscopic knot
8
knot tying
8
diameter measured
8
knot diameter
8
internal diameter
8
diameter mitek
8
+/- 014
8

Similar Publications

It has been reported that the effectiveness of arthroscopic procedures in terms of preventing recurrent dislocation has not matched that from open techniques. Also little is known about how these knots behave when exposed to physiological loading following repair. This study presents the development of a practical tool to allow surgeons to test the quality of their arthroscopic knots and allow them to make choices with regard to knot configuration and suture material.

View Article and Find Full Text PDF

Arthrosopic knot pushers. Does one size fit all?

Int J Clin Pract

May 2008

Oswestry Rotational, Baschurch, Shropshire, UK.

Little is known about how knot-pusher design affects arthroscopic knot tying. In our practice, we observed the knot-pusher riding onto the arthroscopic knot at the point of maximum tightening. This can lead to snagging of the knot, which is undesirable as it may lead to loosening of, or damage to the knot.

View Article and Find Full Text PDF

Purpose: Flexor tendon repair strength depends on the suture technique and the suture material used. Configurations that incorporate locking loops prevent sutures from pulling through the tendon but typically fail because of suture breakage. The choice of suture material therefore influences repair strength.

View Article and Find Full Text PDF

The effect of two nonresorbable suture types on the mechanical performance over a metal suture anchor eyelet.

Knee Surg Sports Traumatol Arthrosc

March 2004

Division of Surgery, University of New South Wales, Prince of Wales Hospital, Edmund Blacket Building, High Street, 2031, Randwick, NSW, Australia.

An understanding of the mechanical properties of different suture materials is valuable when selecting the most appropriate suture and repair technique. Sutures should be strong, easy to handle and have high knot security. The introduction of suture anchors adds an additional variable regarding the effect of stress risers over the eyelet.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!