Objective: To measure the imposed power of breathing (imposed work of breathing per minute) associated with spontaneous breathing through an active impedance threshold device and a sham impedance threshold device.

Design: Prospective randomized blinded protocol.

Setting: University medical center.

Patients: Nineteen healthy, normotensive volunteers (10 males, 9 females, age range 20-56 y, mean +/- SD weight 54.8 +/- 7.7 kg for females, 84 +/- 8 kg for males).

Methods: The volunteers completed 2 trials of breathing through a face mask fitted with an active impedance threshold device set to open at -7 cm H(2)O pressure, or with a sham impedance threshold device, which was identical to the active device except that it did not contain an inspiratory threshold pressure valve diaphragm. Spontaneous breathing frequency (f), tidal volume (V(T)), exhaled minute ventilation, inspiratory pressure, and inspiratory time were measured with a respiratory monitor, and the data were directed to a laptop computer for real-time calculation of the imposed power of breathing.

Results: There were no significant differences in heart rate, respiratory rate, tidal volume, and minute ventilation, with and without inspiratory impedance. For the sham and active impedance threshold device groups, respectively, the mean +/- SD imposed power of breathing values were 0.92 +/- 0.63 J/min and 8.18 +/- 4.52 J/min (p < 0.001), the mean +/- SD inspiratory times were 1.98 +/- 0.86 s and 2.97 +/- 1.1 s (p = 0.001), and the mean +/- SD inspiratory airway/mouth pressures were -1.1 +/- 0.6 cm H(2)O and -11.7 +/- 2.4 cm H(2)O (p < 0.001).

Conclusions: Breathing through an active impedance threshold device requires significantly more power than breathing through a sham device. All subjects tolerated the respiratory work load and were able to complete the study protocol.

Download full-text PDF

Source

Publication Analysis

Top Keywords

impedance threshold
28
threshold device
24
imposed power
16
power breathing
16
active impedance
16
+/-
12
breathing
9
impedance
8
threshold
8
device
8

Similar Publications

Our study evaluated the efficacy and feasibility of left bundle branch area pacing (LBBAP) compared to right ventricular outflow tract septal pacing (RVOSP). We conducted a prospective, single-center, observational study involving 200 consecutive patients who required pacemaker implantation. The patients were divided into two groups (LBBAP and RVOSP), with 100 patients in each group.

View Article and Find Full Text PDF

Objective: High-density nerve cuffs have been successfully utilized to restore somatosensation in individuals with lower-limb loss by interfacing directly with the peripheral nervous system. Elicited sensations via these devices have improved various functional outcomes, including standing balance, walking symmetry, and navigating complex terrains. Deploying neural interfaces in the lower limbs of individuals with limb loss presents unique challenges, particularly due to repetitive muscle contractions and the natural range of motion in the knee and hip joints for transtibial and transfemoral amputees, respectively.

View Article and Find Full Text PDF

Jagim, AR, Horswill, CA, Dobbs, WC, Twohey, EE, Tinsley, GM, Fields, JB, and Jones, MT. Minimum wrestling weight for high school girls wrestling: Time to revisit minimal body fat percent. J Strength Cond Res XX(X): 000-000, 2024-The purpose of this study was to determine whether an alternative body fat percentage (BF%) could be established as a safer margin for minimal wrestling weight (MWW) determination by evaluating the distribution of BF% and MWW values naturally occurring for high school girls wrestling.

View Article and Find Full Text PDF

Paravalvular positioning of a cardiac implantable electronic device electrode during tricuspid valve replacement.

Kardiochir Torakochirurgia Pol

December 2024

Department of Cardiovascular Surgery, Faculty of Medicine, Hacettepe University, Ankara, Turkey.

Introduction: Patients who need tricuspid valve replacement (TVR) surgery often have permanent transvalvular pacemaker (PM) leads, which pose an important challenge in lead management.

Aim: The objective of this study was to evaluate the results of paravalvular positioning of a permanent pacemaker lead during TVR surgery.

Material And Methods: Between 2014 and 2024, a total of 15 patients who had previously had a transvenous pacemaker system underwent TVR.

View Article and Find Full Text PDF

Background: Left bundle branch area pacing (LBBAP) is a new technique for patients with atrioventricular block (AVB) and preserved left ventricular ejection fraction (LVEF), potentially offering better cardiac function than right ventricular pacing (RVP).

Methods: We searched databases and registries for studies that compared LBBAP with RVP in patients with AVB and preserved LVEF. We extracted data on various outcomes and pooled the effect estimates using random-effects models.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!