Obtaining the expertise to perform minimally invasive vascular interventions requires thorough training. In this paper, an algorithm for simulating minimally invasive vascular interventions for training purposes is presented and evaluated. The algorithm enables the simulation of completely straight guide wires as well as intrinsically curved ones based on applied translations and rotations. Friction between the guide wire and the vasculature is incorporated in the model. Quantitative validation is performed by comparing the simulated guide-wire position with the actual position as assessed by 3-D rotational X-ray imaging in physical experiments on a variety of vascular phantoms that truthfully represent human anatomy. The results show that for proper settings of the model's parameters, accurate simulations of guide-wire motion can be obtained, with an average precision of the guide-wire position of around 1.0 mm.

Download full-text PDF

Source
http://dx.doi.org/10.1109/TBME.2006.886659DOI Listing

Publication Analysis

Top Keywords

minimally invasive
12
invasive vascular
12
vascular interventions
12
guide wires
8
guide-wire position
8
modeling friction
4
friction intrinsic
4
intrinsic curvature
4
curvature rotation
4
rotation guide
4

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!