The viability of using polydimethylsiloxane (PDMS) as a retaining material inside a programmed temperature vaporizer injector for the introduction of large-volume samples in gas chromatography is assessed. To that end, materials made up of Volaspher A-2 and coated with different percentages of PDMS (5%, 15%, and 50%) are considered. In addition, adsorbent (Tenax TA) and absorbent (PDMS) materials are comparatively studied in terms of their retention capacity. A relative standard deviation lower than 5.0% is obtained from the injection of PDMS, whereas values up to 49% are provided by Tenax TA. Significantly higher amounts of different volatile compounds are retained by PDMS in comparison with Tenax TA. In conclusion, the use of PDMS as a packing material seems to be viable for large-volume sampling and particularly recommendable for minor compounds occurring in complex matrices.
Download full-text PDF |
Source |
---|---|
http://dx.doi.org/10.1093/chromsci/45.1.33 | DOI Listing |
Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!