A liquid chromatography/electrospray ionization tandem mass spectrometry (LC/ESI-MS/MS) method for measurement of aflatoxins B1, B2, G1, and G2 in maize is described. Aflatoxins (AFs) were extracted from 1 g samples by using tri-portions of acetonitrile/water (80:20, v/v) (10 + 7 + 7 mL), and 2/5 of the extract diluted to 500 mL by water was cleaned up with a 100 mg Carbograph-4 cartridge. After the addition of the internal standard AFM1, the final extract was analyzed by LC/ESI-MS/MS in positive ion mode using multiple reaction monitoring with a triple-quadrupole instrument. A C(18) column thermostatted at 45 degrees C with a mobile phase gradient of acetonitrile/water with 2 mmol/L ammonium formate was used. Although the matrix suppression effect was negligible, quantitation was achieved by an external calibration procedure using matrix-matched standard solutions to improve accuracy. Sample recoveries at four spiking levels ranged from 81 to 101% (relative standard deviation (RSD)