Severity: Warning
Message: file_get_contents(https://...@pubfacts.com&api_key=b8daa3ad693db53b1410957c26c9a51b4908&a=1): Failed to open stream: HTTP request failed! HTTP/1.1 429 Too Many Requests
Filename: helpers/my_audit_helper.php
Line Number: 176
Backtrace:
File: /var/www/html/application/helpers/my_audit_helper.php
Line: 176
Function: file_get_contents
File: /var/www/html/application/helpers/my_audit_helper.php
Line: 250
Function: simplexml_load_file_from_url
File: /var/www/html/application/helpers/my_audit_helper.php
Line: 3122
Function: getPubMedXML
File: /var/www/html/application/controllers/Detail.php
Line: 575
Function: pubMedSearch_Global
File: /var/www/html/application/controllers/Detail.php
Line: 489
Function: pubMedGetRelatedKeyword
File: /var/www/html/index.php
Line: 316
Function: require_once
Purpose: To compare optical low-coherence reflectometry (OLCR) and ultrasound pachymetry in measuring corneal graft thickness in patients after keratoplasty.
Methods: We retrospectively measured the central graft thickness in 41 eyes of 41 patients with the OLCR pachymeter (Haag Streit, Koeniz, Switzerland) and the SP-2000 contact ultrasound pachymeter (Tomey, Nagoya, Japan). Five separate measurements were performed on each eye with both methods. Mean, SD, repeatability, and coefficient of variation of measurements were calculated, and the correlation between the 2 methods was studied with Spearman regression.
Results: Mean central graft thickness was 546 +/- 51 (SD) microm with the contact ultrasound pachymeter and 546 +/- 47 microm with the OLCR pachymeter. The correlation between both methods was strong (rs = 0.96). No significant differences in mean SD of measurements were observed between OLCR pachymetry (mean SD = 4.66 microm) and contact ultrasound pachymetry (mean SD = 4.88 microm). The repeatability of both methods was comparable (P = 0.06) and high (the average coefficient of variation of the central corneal graft thickness was 0.9% with both pachymeters). The postoperative time did not affect the correlation between both pachymeters (P > 0.05).
Conclusions: Central corneal graft thickness values obtained with the OLCR pachymeter were similar to those obtained with a contact ultrasound pachymeter. In some cases of lamellar keratoplasty, the corneal refractive index could change at the interface level that could affect OLCR measurements.