Backside damage corresponding to articular damage in retrieved tibial polyethylene inserts.

Clin Orthop Relat Res

Orthopaedic Research Laboratory, The BioMotion Foundation, Palm Beach, FL 33480-0248, USA.

Published: May 2007

AI Article Synopsis

  • The study investigates the causes of backside wear in polyethylene inserts used in tibial components, suggesting that both motion between the insert and tibial baseplate, as well as articular contact stresses, play a role.
  • 37 tibial inserts were analyzed, revealing that the extent of damage was influenced by the duration of physiologic loading and demonstrated a mechanical interlock between the insert and tibial tray.
  • Findings indicate that inserts with less motion experienced greater wear and longer functional lifespan, with damage patterns showing significant differences between articular and backside interfaces, implying distinct wear mechanisms.

Article Abstract

Although motion between the polyethylene insert and tibial baseplate is one probable cause of backside wear, articular contact stresses and kinematic conditions may be additional factors. However, comparisons of articular and backside damage patterns are limited. We report the effect of physiologic loading on the modular capture mechanism and distribution of articular and backside surface damage patterns on retrieved tibial components. We evaluated damage patterns on 37 tibial inserts with a full peripheral rim capture mechanism, including six autopsy-retrieved components that were not previously disassembled and were available for mechanical testing. The duration of physiologic loading affected the modular capture mechanism and damage patterns. Backside damage revealed evidence of a mechanical interlock between the polyethylene insert and tibial tray consistent with the measured insert motion. In autopsy components retrieved after 2 to 6 years, inserts with the least motion had the longest duration of in vivo function and the largest backside damage area. The backside damage area and location corresponded to articular damage with damage patterns concentrated on the posterior half of the polyethylene inserts. Substantial differences between the articular and backside damage modes suggest different wear mechanisms exist at the two interfaces during physiologic loading.

Download full-text PDF

Source
http://dx.doi.org/10.1097/BLO.0b013e3180320b01DOI Listing

Publication Analysis

Top Keywords

backside damage
24
damage patterns
20
articular backside
12
physiologic loading
12
capture mechanism
12
damage
11
backside
8
articular damage
8
retrieved tibial
8
polyethylene inserts
8

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!