Prevalence and characterization of asymptomatic pacemaker and ICD lead perforation on CT.

Pacing Clin Electrophysiol

Department of Radiology, Division of Cardiology, Albert Einstein College of Medicine, Montefiore Medical Center, Bronx, New York, USA.

Published: January 2007

Background: Pacemakers and implantable cardiac defibrillators (ICDs) are widely used for the management of cardiac arrhythmias and congestive heart failure (CHF). Acute implantation complication rates range from 3% to 7%. The aim of this study is to describe the incidence of lead perforation on computed tomography (CT), and correlate these findings with electrophysiologic data.

Methods: Images of 100 consecutive patients with permanent pacemakers (n = 72) or ICDs (n = 28) who underwent multidetector CTs of the chest were identified. Cases were reviewed by 2 cardiothoracic radiologists, and a third if there was disagreement. Each CT was reviewed for device and fixation type, tip position, and presence of pericardial effusion. Results were correlated with lead impedance and pacing threshold, when available [79% (79/100)]. A cardiac electrophysiologist interpreted device data.

Results: All 100 patients had right ventricular leads (58 passive, 42 active) and 61 had right atrial leads (12 passive, 49 active). 15% (15/100) of patients had a lead perforation. Perforation rates were 15% (9/61) for atrial and 6% (6/100) for ventricular leads (P < 0.05, chi square). Four of 28 (14%) right ventricular ICD leads and 2 of 72 (3%)right ventricular pacemaker leads were perforated (P < 0.05, chi square). 12% (6/49) of active right atrial leads, and 25% (3/12) of passive right atrial leads perforated (P = NS, chi square). 7% (3/42) of active right ventricular leads, and 5% (3/58) of passive ventricular leads perforated (P = NS, chi square). Electrophysiologic parameters did not differ significantly between perforated and nonperforated leads.

Conclusion: Asymptomatic perforation is a common phenomenon and rarely resulting in electrophysiologic consequences. Atrial leads perforated more frequently than ventricular leads, and ventricular ICD leads perforated more frequently than ventricular pacemaker leads.

Download full-text PDF

Source
http://dx.doi.org/10.1111/j.1540-8159.2007.00575.xDOI Listing

Publication Analysis

Top Keywords

ventricular leads
20
leads perforated
20
atrial leads
16
chi square
16
leads
13
lead perforation
12
ventricular
9
leads passive
8
passive active
8
active atrial
8

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!