Background: Retinal implants are intended to replace photoreceptors in patients suffering from degenerative retinal diseases such as retinitis pigmentosa. Data show that photodiodes in subretinal implants are not powerful enough to stimulate overlying retinal tissue by simply transforming light energy into electrical energy. Therefore, infrared (IR) irradiation has been envisioned to supply additional energy. While epiretinal implants mostly use induction coils for wireless energy transfer, IR irradiation seems to be an additional option. This study investigated the feasibility of an IR energy supply for an active subretinal implant by assessing thermal effects of IR irradiation onto the rabbit retina.
Methods: Polyimide foil strips carrying an optical sensor as well as a thermal sensor were implanted into the subretinal space of the eyes of nine rabbits using a transchoroidal surgical approach. The area of the thermal sensor was irradiated by an IR laser (830 nm) focused on the device. The sensor provided simultaneous real-time measurements of absolute temperature and irradiation density, allowing direct correlation of the temperature increase to different intensities of IR irradiation. Possible IR-related damage to the retina was examined in histological sections. Temperature changes in living and dead animals were evaluated as a function of IR irradiation power of between 0.1 mW and 40 mW (0.03 mW/mm2-12.7 mW/mm2).
Results: We found an exponential relationship between IR irradiation power and temperature increase over the whole range (up to 12.7 mW/mm2) in the living animal. The maximum temperature increase caused by IR irradiation of 40 mW (12.7 mW/mm2) was 4.5 degrees C. The ratio of temperature increase to IR irradiation density postmortem (i.e., without ocular blood flow) was linear over the whole range, with 1.15 degrees C per 1 mW/mm2. Thus, the cooling effect of ocular blood flow varied depending on IR irradiance density. In histological sections, no IR-induced damage to the retina was detected.
Conclusions: A temperature increase of 3.2 degrees C in the living rabbit eye is to be expected when powering a subretinal implant with 15 mW (4.8 mW/mm2) IR power, the wattage used in an external power supply for an active implant with 1,500 electrodes. This appears to be a tolerable increase for ocular tissue.
Download full-text PDF |
Source |
---|---|
http://dx.doi.org/10.1007/s00417-006-0513-1 | DOI Listing |
Appl Biochem Biotechnol
January 2025
Graduate School of Pharmaceutical Sciences, Osaka University, Suita Yamadaoka 1-6, Suita, Osaka, 565-0871, Japan.
γ-Glutamylcysteine (γ-EC) can increase intracellular glutathione (GSH) levels, which may prevent and alleviate age-related disorders and chronic diseases caused by oxidative damage. However, the commercial availability of γ-EC remains limited owing to its complex chemical synthesis from glutamate and cysteine. In this study, we have developed the method of the effective conversion of GSH to γ-EC to achieve the optimal reaction conditions for repeated batch production and potential application in industrial γ-EC production using the phytochelatin synthase-like enzyme NsPCS.
View Article and Find Full Text PDFEnviron Sci Pollut Res Int
January 2025
Department of Chemistry, Banasthali Vidhyapith, Banasthali, Rajasthan, 304022, India.
Plant extracts and bacterial biofilm are acknowledged to offer impressive corrosion-inhibitory activities. However, anticorrosive properties of their combination are still less reported. Thus, in the present study, we aimed to evaluate the corrosion inhibition efficiency of Saccharum officinarum bagasse (SOB) plant extract, Pseudomonas chlororaphis (P.
View Article and Find Full Text PDFJ Vector Borne Dis
January 2025
Department of Internal Medicine (ID Division), All India Institute of Medical Sciences, Rishikesh, Uttarakhand, India.
Dengue hemorrhagic fever (DHF) typically presents with various bleeding manifestations such as epistaxis, gum bleeding, and gastrointestinal bleeding. However, spontaneous large muscle hematoma formation is a rare complication. This case report discusses a patient with DHF who developed bilateral psoas muscle hematomas, a very uncommon presentation.
View Article and Find Full Text PDFJ Vector Borne Dis
January 2025
ICMR-Vector Control Research Centre Field Station, Kottayam, Kerala, India.
West Nile virus (WNV) infection is emerging as a disease of public health concern in Kerala, India with recurring outbreaks since 2011. With its tropical climate, biodiversity hot spots of Western ghats, forest cover, plenty of water bodies and bird sanctuaries, Kerala provides an ideal ecological niche for vector breeding and transmission of WNV. In this article, we reflect on the peculiar features of the outbreaks of WNV fever in Kerala and highlight the knowledge gaps, the research priorities and the need for effective control measures.
View Article and Find Full Text PDFJ Vector Borne Dis
January 2025
State Virology Laboratory, Department of Microbiology, Gandhi Medical College, Bhopal, Madhya Pradesh, India.
Background Objectives: Co-infection of dengue virus and acute hepatitis A virus in paediatric population is a major health concern in endemic countries. This cross sectional retrospective study was conducted to evaluate the prevalence of hepatitis A virus among the clinically dengue suspected paediatric cases presented at our tertiary care centre during the two-year period (2022-2023).
Methods: A total of 747 dengue suspected paediatric clinical specimens were included in this study.
Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!