[Prospective comparison of different ultrasound modalities to measure thicknesses less than 1 mm].

Rofo

Kinderradiologie, Vestische Kinderklinik, Datteln.

Published: January 2007

AI Article Synopsis

Article Abstract

Purpose: New ultrasound techniques with improved visualization and better transducers constantly expand the methods of measurement. In the border area in the case of thickness measurements of less than 1 millimeter, e. g. the measurement of the "Intima-Media Thickness" (IMT), it is not clear if better visualization comes at the price of incorrect measurement values. We checked different modalities for accuracy and visualization quality.

Materials And Methods: We prospectively examined the IMT of obese children using different ultrasound modalities. The image quality was evaluated by two independent pediatric radiologists. A phantom with an exact thickness of 0.3 mm (value of the thinnest measured IMT thickness) was subsequently created by precision engineers. Measurements were carried out and analyzed using fundamental ultrasound, tissue harmonic imaging, speckle reduction imaging and cross beam (real-time compound) imaging.

Results: The best image quality was reached using speckle reduction imaging. This technique and fundamental ultrasound produced the most exact data. However, the recognizability was worse in fundamental ultrasound than in speckle reduction imaging.

Conclusion: Speckle reduction imaging appears to be the best technique for measuring thicknesses less than 1 mm.

Download full-text PDF

Source
http://dx.doi.org/10.1055/s-2006-927237DOI Listing

Publication Analysis

Top Keywords

speckle reduction
16
fundamental ultrasound
12
reduction imaging
12
ultrasound modalities
8
image quality
8
ultrasound
6
[prospective comparison
4
comparison ultrasound
4
modalities measure
4
measure thicknesses
4

Similar Publications

The angular dependence of random laser (RL) generation in a commercially available rhodamine 6G (Rh6G) dye has been demonstrated using ZnO-CuO heterostructure as passive scatterers. The grass-like superstructure formed at a 1M:1M molar ratio of ZnO-CuO significantly enhances scattering, resulting in RL spikes with a full width at half maximum (FWHM) of just a few nanometer and a noticeable reduction in the RL threshold. RL emission spectra were collected over an angular spread of 0-180 degrees, revealing a remarkable shift in RL emission from 566 nm to 580 nm.

View Article and Find Full Text PDF

Objectives: Supra-normal left ventricular ejection fraction (snLVEF) represents a heterogeneous group with distinct prognoses. Left atrial (LA) strain, measured by speckle tracking echocardiography (STE), is a validated prognostic indicator. This study aimed to evaluate LA and left ventricular (LV) mechanical strains in hypertensive patients with snLVEF.

View Article and Find Full Text PDF

Abnormal tau phosphorylation is a key mechanism in neurodegenerative diseases. Evidence implicates infectious agents, such as Herpes Simplex Virus 1 (HSV-1), as co-factors in the onset or the progression of neurodegenerative diseases, including Alzheimer's disease. This has led to divergence in the field regarding the contribution of viruses in the etiology of neurodegenerative diseases.

View Article and Find Full Text PDF

Spectrally encoded flow cytometry using few-mode fiber collection.

Biomed Opt Express

January 2025

Faculty of Biomedical Engineering, Technion-Israel Institute of Technology, Haifa 3200003, Israel.

In fiber-based confocal microscopy, using two separate fibers for illumination and collection enables the use of a few-mode fiber to achieve an effect similar to opening the pinhole in a conventional confocal microscope. In some Fourier-domain applications, however, or when a spectral measurement is involved, the coherent light detection would lead to noticeable spectral modulation artifacts that result from differential mode delay, an effect caused by the multimode propagation in the collection fiber. After eliminating these artifacts by using mode-dependent polarization control, we demonstrate effective spectrally encoded imaging with improved signal efficiency and lower speckle noise, and only a minor, negligible reduction in lateral and axial resolutions.

View Article and Find Full Text PDF

Background: Hepatocellular carcinoma (HCC) is characterized by high postoperative recurrence rates, and predicting early recurrence is crucial for improving clinical outcomes, yet remains challenging. Both preoperative computed tomography (CT) imaging radiomic features and serum biomarkers related to microvascular infiltration are important indicators of HCC prognosis. This study aimed to develop a nomogram model incorporating both preoperative CT radiomic features and serum biomarkers associated with microvascular infiltration to predict early postoperative recurrence in HCC patients.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!